CLASS II ONGOING Device recall · Reported 17 Dec 2025 · FDA Enforcement Report

The Amsco 7052HP Washer/Disinfector is intended for use recalled

Steris is recalling The Amsco 7052HP Washer/Disinfector is intended for use in the cleaning and, distributed nationwide in the US. The recall was initiated on 19 Nov 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: UDI:00724995176587, 00724995176594, 00724995176570 Serial number of distributed units: 3619625011, 3621625009, 3621625012, 3621925006, 3621925009, 3622425006, 3622725007, 3623425005, 3623825002, 3624525002
Quantity recalled10

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0909-2026
FDA event ID
97986
Recalling firm
Steris, Mentor, OH
Classification
Class II
Status
Ongoing
Recall initiated
19 Nov 2025
FDA classified
11 Dec 2025
Reported
17 Dec 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

19 Nov 2025Recall initiated by company
11 Dec 2025Classified by FDA+22 days
17 Dec 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.

Where it was distributed

Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.

Nationwide ArizonaColoradoFloridaGeorgiaIllinoisKentuckyLouisianaMichiganNorth CarolinaNew HampshireNew YorkPennsylvaniaTennesseeTexasUtahVirginiaWashingtonWisconsin

Questions about this recall

Why was it recalled?

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

Which lots are affected?

Lot Code: UDI:00724995176587, 00724995176594, 00724995176570 Serial number of distributed units: 3619625011, 3621625009, 3621625012, 3621925006, 3621925009, 3622425006, 3622725007, 3623425005, 3623825002, 3624525002

Where was it sold?

Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 97986
ReportedClassRecallCompanyReasonWhere
CLASS IIThe Amsco 7053HP Washer/Disinfector is intended for use recalledSterisForeign MaterialNationwide

More recalls from Steris

All 113

Other Surgical Instruments recalls

All 3,809