CLASS II ONGOING Device recall · Reported 2 Apr 2025 · FDA Enforcement Report
Checkpoint Guardian Intraoperative Lead, Medium recalled
Checkpoint Surgical is recalling Checkpoint Guardian Intraoperative Lead, Medium, distributed nationwide in the US. The recall was initiated on 24 Feb 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1467-2025
- FDA event ID
- 96387
- Recalling firm
- Checkpoint Surgical, Cleveland, OH
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 24 Feb 2025
- FDA classified
- 26 Mar 2025
- Reported
- 2 Apr 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Visit
- Reason category
- Foreign Material
- Device type
- Surgical Instruments
Timeline
| 24 Feb 2025 | Recall initiated by company | |
| 26 Mar 2025 | Classified by FDA | +30 days |
| 2 Apr 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Medium Model/Catalog Number: REF# 9525 (Medium) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.
Where it was distributed
US Nationwide distribution.
Questions about this recall
Why was it recalled?
Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.
Which lots are affected?
UDI-DI B13095250 Medium Lot 4322 Expires 01Apr2026 Lot 4326 Expires 01Apr2026 Lot 4333 Expires 01Jul2026 Lot 4352 Expires 01Jul2026
Where was it sold?
US Nationwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 96387This product is part of a recall event; the main page for the event is Checkpoint Guardian Intraoperative Lead, Small recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Checkpoint Guardian Intraoperative Lead, Small recalled | Checkpoint Surgical | Foreign Material | Nationwide |
More recalls from Checkpoint Surgical
All 2| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Checkpoint Guardian Intraoperative Lead, Small recalled | Checkpoint Surgical | Foreign Material | Nationwide |
Other Surgical Instruments recalls
All 3,921| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Sterile, synthetic, absorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | TruSystem 7500 surgical table recalled by Baxter Healthcare | Baxter Healthcare | Other | Nationwide |
| Device | CLASS II | Sterile, natural, nonabsorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | Sterile, natural, nonabsorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | Sterile, non-absorbable 7-0 blue polypropylene surgical recalled | Henry Schein | Other | Nationwide |