CLASS II Device recall · Reported 19 Sep 2012 · FDA Enforcement Report

Enduramesh Corpectomy Spacer System recalled by Lucero Medical

Lucero Medical is recalling Enduramesh Corpectomy Spacer System, distributed nationwide in the US. The recall was initiated on 15 Jul 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart #-Lot # (s): 762.010L-110738; 762.013L-110738; 762.017L-60611, 90611, and 512171; 762.019L-60611, 90611, 100308, 1261901, and 20121901; 762.021L-60611, 90611, 100308, 110106, 20122101, and 20122102; 762.023L- 90611, 100209, 100218, 00706, 10709, 20122301, and 20122302; 762.025L-90611, 100218, 110106, 20122501, and 20122502; 762.027L-60611, 90611, 100706, and 20122071; 762.029L-90611, 1262901, and 90611; 762.031L-1263101; 762.033L-90611, 91002, and 1263301; 762.041L-90611 and 1264101; 762.043L-90611 and 1264301; 762.045L-90611; 762.047L-90611; 762.049L-90611; 762.051L-90611 and 512511; 762.053L-90611; 762.090L-90611; 762.090L-2.0-100206 and 100342; 762.090L-2.5-100206; 762.159L-110505; 762.237L-91002; 762.243L-91018; 762.270L-91018; 762.290L-91018; 762.731L-91018; 762.735L-91002; 762.767L-91018; and 762.790L-91018.
Quantity recalled169

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2368-2012
FDA event ID
62926
Recalling firm
Lucero Medical, Independence, OH
Classification
Class II
Status
Terminated
Recall initiated
15 Jul 2012
FDA classified
11 Sep 2012
Reported
19 Sep 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility
Device type
Spinal Implants

Timeline

15 Jul 2012Recall initiated by company
11 Sep 2012Classified by FDA+58 days
19 Sep 2012Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass sterilization validation.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Enduramesh Corpectomy Spacer System Product Usage: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine (T1-S1) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma/fracture. The device is intended for use with either allograft or autograft. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference.

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass sterilization validation.

Which lots are affected?

Part #-Lot # (s): 762.010L-110738; 762.013L-110738; 762.017L-60611, 90611, and 512171; 762.019L-60611, 90611, 100308, 1261901, and 20121901; 762.021L-60611, 90611, 100308, 110106, 20122101, and 20122102; 762.023L- 90611, 100209, 100218, 00706, 10709, 20122301, and 20122302; 762.025L-90611, 100218, 110106, 20122501, and 20122502; 762.027L-60611, 90611, 100706, and 20122071; 762.029L-90611, 1262901, and 90611; 762.031L-1263101; 762.033L-90611, 91002, and 1263301; 762.041L-90611 and 1264101; 762.043L-90611 and 1264301; 762.045L-90611; 762.047L-90611; 762.049L-90611; 762.051L-90611 and 512511; 762.053L-90611; 762.090L-90611; 762.090L-2.0-100206 and 100342; 762.090L-2.5-100206; 762.159L-110505; 762.237L-91002; 762.243L-91018; 762.270L-91018; 762.290L-91018; 762.731L-91018; 762.735L-91002; 762.767L-91018; and 762.790L-91018.

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other Spinal Implants recalls

All 1,157