CLASS II ONGOING Device recall · Reported 2 Sep 2020 · FDA Enforcement Report
Flexible Intubation Video Endoscope, Part: 11301BNX recalled
Karl Storz Endoscopy is recalling Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US, distributed nationwide in the US. The recall was initiated on 27 Jul 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2891-2020
- FDA event ID
- 86009
- Recalling firm
- Karl Storz Endoscopy, El Segundo, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 27 Jul 2020
- FDA classified
- 27 Aug 2020
- Reported
- 2 Sep 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Surgical Instruments
Timeline
| 27 Jul 2020 | Recall initiated by company | |
| 27 Aug 2020 | Classified by FDA | +31 days |
| 2 Sep 2020 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Where it was distributed
Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Is Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US recalled?
Yes. Karl Storz Endoscopy recalled Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US on 27 Jul 2020. The recall is Class II and its status is Ongoing. Recall number Z-2891-2020.
Why was it recalled?
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
Which lots are affected?
Instruction Manual Version: 01/2020
Where was it sold?
Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Karl Storz Endoscopy
All 71| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Flexible Hystero-Fiberscope Kits recalled by Karl Storz Endoscopy | Karl Storz Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Karl Storz - Endoskope recalled | Karl Storz Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Karl Storz - Endoskope recalled | Karl Storz Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Karl Storz - Endoskope recalled | Karl Storz Endoscopy | Other | Nationwide |
| Device | CLASS II | Karl Storz - Endoskope recalled | Karl Storz Endoscopy | Sterility | Nationwide |
Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Sterility | Nationwide |