CLASS II ONGOING Device recall · Reported 2 Sep 2020 · FDA Enforcement Report

Flexible Intubation Video Endoscope, Part: 11301BNX recalled

Karl Storz Endoscopy is recalling Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US, distributed nationwide in the US. The recall was initiated on 27 Jul 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesInstruction Manual Version: 01/2020
Quantity recalled2083

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2891-2020
FDA event ID
86009
Recalling firm
Karl Storz Endoscopy, El Segundo, CA
Classification
Class II
Status
Ongoing
Recall initiated
27 Jul 2020
FDA classified
27 Aug 2020
Reported
2 Sep 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

27 Jul 2020Recall initiated by company
27 Aug 2020Classified by FDA+31 days
2 Sep 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.

Where it was distributed

Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Is Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US recalled?

Yes. Karl Storz Endoscopy recalled Flexible Intubation Video Endoscope, Part: 11301BNX, with Instruction Manual: Z18444US on 27 Jul 2020. The recall is Class II and its status is Ongoing. Recall number Z-2891-2020.

Why was it recalled?

Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.

Which lots are affected?

Instruction Manual Version: 01/2020

Where was it sold?

Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Karl Storz Endoscopy

All 71

Other Surgical Instruments recalls

All 3,809