CLASS II ONGOING Device recall · Reported 18 Jan 2023 · FDA Enforcement Report
Large Rotating Craniotome Attachment recalled by The Anspach Effort
The Anspach Effort is recalling Large Rotating Craniotome Attachment, distributed nationwide in the US. The recall was initiated on 7 Dec 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0922-2023
- FDA event ID
- 91231
- Recalling firm
- The Anspach Effort, Palm Beach Gardens, FL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 Dec 2022
- FDA classified
- 6 Jan 2023
- Reported
- 18 Jan 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Processing Deviation
- Device type
- Spinal Implants
Timeline
| 7 Dec 2022 | Recall initiated by company | |
| 6 Jan 2023 | Classified by FDA | +30 days |
| 18 Jan 2023 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.
Product as listed by the FDA
7.5 cm Large Rotating Craniotome Attachment. Use with XMax¿, microMax, and eMax¿ Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
Where it was distributed
US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania Macedonia Mauritius Morocco Netherlands Poland Portugal Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom of Great Britain and Northern Ireland Argentina Bolivia Brazil Chile Colombia Costa Rica Jamaica Mexico Panama Peru Canada
Questions about this recall
Is Large Rotating Craniotome Attachment recalled?
Yes. The Anspach Effort recalled Large Rotating Craniotome Attachment on 7 Dec 2022. The recall is Class II and its status is Ongoing. Recall number Z-0922-2023.
Why was it recalled?
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Which lots are affected?
GTIN: 00845384001768 All lots
Where was it sold?
US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania Macedonia Mauritius Morocco Netherlands Poland Portugal Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom of Great Britain and Northern Ireland Argentina Bolivia Brazil Chile Colombia Costa Rica Jamaica Mexico Panama Peru Canada
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from The Anspach Effort
All 260| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Adult Craniotome recalled by The Anspach Effort | The Anspach Effort | Processing Deviation | Nationwide |
| Device | CLASS II | Adult Rotating Craniotome Attachment recalled by The Anspach Effort | The Anspach Effort | Processing Deviation | Nationwide |
| Device | CLASS II | Adult Craniotome, Thin Foot Plate Attachment recalled | The Anspach Effort | Processing Deviation | Nationwide |
| Device | CLASS II | Adult Rotating Craniotome Attachment recalled by The Anspach Effort | The Anspach Effort | Processing Deviation | Nationwide |
| Device | CLASS II | Large Craniotome Attachment Use with XMax¿, microMax recalled | The Anspach Effort | Processing Deviation | Nationwide |
Other Spinal Implants recalls
All 1,157| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | AVID Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: ENT recalled | AVID Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: A-P PACK recalled | AVID Medical | Sterility | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |