CLASS II ONGOING Device recall · Reported 18 Jan 2023 · FDA Enforcement Report

Large Rotating Craniotome Attachment recalled by The Anspach Effort

The Anspach Effort is recalling Large Rotating Craniotome Attachment, distributed nationwide in the US. The recall was initiated on 7 Dec 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGTIN: 00845384001768 All lots
Quantity recalled6059 units in total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0922-2023
FDA event ID
91231
Recalling firm
The Anspach Effort, Palm Beach Gardens, FL
Classification
Class II
Status
Ongoing
Recall initiated
7 Dec 2022
FDA classified
6 Jan 2023
Reported
18 Jan 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Processing Deviation
Device type
Spinal Implants

Timeline

7 Dec 2022Recall initiated by company
6 Jan 2023Classified by FDA+30 days
18 Jan 2023Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.

Product as listed by the FDA

7.5 cm Large Rotating Craniotome Attachment. Use with XMax¿, microMax, and eMax¿ Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium

Where it was distributed

US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania Macedonia Mauritius Morocco Netherlands Poland Portugal Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom of Great Britain and Northern Ireland Argentina Bolivia Brazil Chile Colombia Costa Rica Jamaica Mexico Panama Peru Canada

Nationwide Georgia

Questions about this recall

Is Large Rotating Craniotome Attachment recalled?

Yes. The Anspach Effort recalled Large Rotating Craniotome Attachment on 7 Dec 2022. The recall is Class II and its status is Ongoing. Recall number Z-0922-2023.

Why was it recalled?

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

Which lots are affected?

GTIN: 00845384001768 All lots

Where was it sold?

US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finland France & French Guiana Georgia Germany Greece Hungary Israel Italy Kenya Kuwait Lebanon Lithuania Macedonia Mauritius Morocco Netherlands Poland Portugal Saudi Arabia Slovakia South Africa Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom of Great Britain and Northern Ireland Argentina Bolivia Brazil Chile Colombia Costa Rica Jamaica Mexico Panama Peru Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from The Anspach Effort

All 260

Other Spinal Implants recalls

All 1,157