CLASS I ONGOING Device recall · Reported 23 Aug 2023 · FDA Enforcement Report
Mahurkar High Pressure Triple Lumen Acute Dialysis Catheter recalled
Covidien is recalling Mahurkar High Pressure Triple Lumen Acute Dialysis Catheter, Curved Extensions, distributed nationwide in the US. The recall was initiated on 28 Jun 2023 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2382-2023
- FDA event ID
- 92660
- Recalling firm
- Covidien, Mansfield, MA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 28 Jun 2023
- FDA classified
- 11 Aug 2023
- Reported
- 23 Aug 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Particulate Matter
- Device type
- Heart & Vascular Devices
Timeline
| 28 Jun 2023 | Recall initiated by company | |
| 11 Aug 2023 | Classified by FDA | +44 days |
| 23 Aug 2023 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Product as listed by the FDA
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions REF: 8888345512HP
Where it was distributed
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
Questions about this recall
Is Mahurkar High Pressure Triple Lumen Acute Dialysis Catheter, Curved Extensions recalled?
Yes. Covidien recalled Mahurkar High Pressure Triple Lumen Acute Dialysis Catheter, Curved Extensions on 28 Jun 2023. The recall is Class I and its status is Ongoing. Recall number Z-2382-2023.
Why was it recalled?
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Which lots are affected?
GTIN: 20884521128108 LOT Numbers: 2018800014 2023400145 2027200276 2119400215 2130800064 2201000047
Where was it sold?
US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Covidien
All 493| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | EndoFLIP System that includes recalled by Covidien | Covidien | Other | Nationwide |
| Device | CLASS II | Dermalon 88861757-31 Dermalon* 4-0 Blu C13 recalled by Covidien | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Surgipro SP-623 Surgipro* 2-0 BLU SC2 X36 SP recalled by Covidien | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | V-Loc 90 Absorbable Wound Closure Device Vloc 90 0 VL recalled | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Surgipro II S8698GX Surgipro II BL 5/0 18 P-13 S8698GX recalled | Covidien | Lead & Heavy Metals | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |