CLASS II Device recall · Reported 18 Jul 2012 · FDA Enforcement Report
Pathfinder NXT Fixed Percutaneous Rod Holder recalled by Zimmer
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle), distributed nationwide in the US. The recall was initiated on 5 Jun 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1980-2012
- FDA event ID
- 62325
- Recalling firm
- Zimmer, Warsaw, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 5 Jun 2012
- FDA classified
- 10 Jul 2012
- Reported
- 18 Jul 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Spinal Implants
Timeline
| 5 Jun 2012 | Recall initiated by company | |
| 10 Jul 2012 | Classified by FDA | +35 days |
| 18 Jul 2012 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) , Stainless Steel / Silicone Rubber, Rx only, non sterile. Product Usage: The rod is held by the instrument for inserting and guiding into the screw heads via percutaneous delivery during minimally invasive surgery to fuse the spine. Per the surgical technique (Ll554 Pathfinder NXT Surgical Technique);
Where it was distributed
Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
Questions about this recall
Is Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) recalled?
Yes. Zimmer recalled Pathfinder NXT Fixed Percutaneous Rod Holder, Part numbers 3573-1 (angle handle) on 5 Jun 2012. The recall is Class II and its status is Terminated. Recall number Z-1980-2012.
Why was it recalled?
Zimmer is recalling Pathfinder NXT Fixed Percutaneous Rod Holder, because complaints have been received that the rod holder tip may crack or break when the rod is manipulated during surgery.
Which lots are affected?
Part numbers 3573-1 (angle handle) all lots including 64PY, 67YA, 69FF, 71GB, 72JL, 73GY
Where was it sold?
Worldwide Distribution - USA (nationwide) and the countries of: Austria, Australia, Belgium, France, Gelmany, Italy, Spain, Switzerland and the United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Zimmer
All 539| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Off-Axis Comprehensive Shoulder System, Large, Augment recalled | Zimmer | Foreign Material | Nationwide |
| Device | CLASS II | Off-Axis Comprehensive Shoulder System Small, Augment recalled by Zimmer | Zimmer | Foreign Material | Nationwide |
| Device | CLASS II | Off-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide recalled | Zimmer | Foreign Material | Nationwide |
| Device | CLASS II | Persona Revision¿ Trabecular Metal¿ Femoral Distal Augment recalled | Zimmer | Foreign Material | IN |
| Device | CLASS II | Off-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide recalled | Zimmer | Foreign Material | Nationwide |
Other Spinal Implants recalls
All 1,157| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | AVID Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: ENT recalled | AVID Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: A-P PACK recalled | AVID Medical | Sterility | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits, labeled as recalled | Medline Industries | Other | Nationwide |