CLASS II ONGOING Device recall · Reported 5 Nov 2025 · FDA Enforcement Report
Safe-t-fill Micro Capillary Blood Collection recalled by ASP Global
ASP Global is recalling Safe-t-fill Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium Edta, distributed nationwide in the US. The recall was initiated on 5 May 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0372-2026
- FDA event ID
- 97551
- Recalling firm
- ASP Global, Austell, GA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 5 May 2025
- FDA classified
- 27 Oct 2025
- Reported
- 5 Nov 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Syringes & Needles
Timeline
| 5 May 2025 | Recall initiated by company | |
| 27 Oct 2025 | Classified by FDA | +175 days |
| 5 Nov 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910
Where it was distributed
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
Nationwide AlaskaAlabamaArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIdahoIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippi
Questions about this recall
Is Safe-t-fill Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium Edta recalled?
Yes. ASP Global recalled Safe-t-fill Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium Edta on 5 May 2025. The recall is Class II and its status is Ongoing. Recall number Z-0372-2026.
Why was it recalled?
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Which lots are affected?
Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025.
Where was it sold?
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from ASP Global
All 25| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | A clear, plastic capillary blood collection tube coated with Heparin recalled | ASP Global | Foreign Material | Nationwide |
| Device | CLASS II | The Safe-t-fill Capillary Blood Collection System is recalled | ASP Global | Other | Nationwide |
| Device | CLASS II | The Safe-t-fill Capillary Blood Collection System is recalled | ASP Global | Other | Nationwide |
| Device | CLASS II | Safe-t-fill Micro Collection Tube Extender recalled by ASP Global | ASP Global | Other | Nationwide |
| Device | CLASS II | A clear, plastic capillary blood collection tube recalled by ASP Global | ASP Global | Foreign Material | Nationwide |
Other Syringes & Needles recalls
All 1,060| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | BD Needle Kit for Powered Driver recalled by Bard Access Systems | Bard Access Systems | Other | 32 states |
| Device | CLASS II | Boston Scientific Enhance Transcarotid/Peripheral Access Kit recalled | Galt Medical | Other | Nationwide |
| Device | CLASS II | Geneva Brachytherapy Applicator recalled by Elekta | Elekta | Other | Nationwide |
| Device | CLASS II | Brachytherapy Applicator Advanced Gynecological Applicator recalled | Elekta | Other | Nationwide |
| Device | CLASS II | Medline 111 Gravity Set W/3 Needle Free Y-Sites recalled | Bd Switzerland Sarl | Other | IL |