CLASS II ONGOING Device recall · Reported 30 Sep 2020 · FDA Enforcement Report

Speed Stitch Needle Cassette recalled by ArthroCare

ArthroCare is recalling Speed Stitch Needle Cassette, distributed nationwide in the US. The recall was initiated on 17 Aug 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526
Quantity recalled6,366 needles

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2999-2020
FDA event ID
86290
Recalling firm
ArthroCare, Austin, TX
Classification
Class II
Status
Ongoing
Recall initiated
17 Aug 2020
FDA classified
21 Sep 2020
Reported
30 Sep 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

17 Aug 2020Recall initiated by company
21 Sep 2020Classified by FDA+35 days
30 Sep 2020Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Speed Stitch Needle Cassette

Where it was distributed

Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV. International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyMassachusettsMarylandMaine

Questions about this recall

Is Speed Stitch Needle Cassette recalled?

Yes. ArthroCare recalled Speed Stitch Needle Cassette on 17 Aug 2020. The recall is Class II and its status is Ongoing. Recall number Z-2999-2020.

Why was it recalled?

Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.

Which lots are affected?

Model: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526

Where was it sold?

Domestic Distribution: AK, AL, AR, AZ, CA, CO, CT,DB,FL,GA, HI, IA, ID, IL, IN, KS, KY, MA,MD,ME,MI,MN,MO,MS,MT, NC,ND, NE, NJ, NSW, NV,NY, OH,OK, OR, PA,SC, TN,TX, UT, VA, WA, WI, WV. International Distribution: AE, AU, CH, DE, FI, FR, GB, HK, IE, PA, PR, PT, RU, SG, and ZA

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from ArthroCare

All 5
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IISmith&Nephew SmartStitch PerfectPasser Connector recalled by ArthroCareArthroCareOther43 states
DeviceCLASS IITopaz MicroDebrider recalled by ArthroCareArthroCareOtherNationwide
DeviceCLASS IICavity Spine Wand indicated for resection, ablation recalledArthroCareSterilityNationwide
DeviceCLASS IIEVac 70 XTRA with Integrated Cable and PROcise EZ View recalledArthroCareOtherNationwide

Other Syringes & Needles recalls

All 1,060