CLASS II ONGOING Device recall · Reported 28 Oct 2026 · FDA Enforcement Report

Techno-Path Multichem AE Unassayed Level recalled

Techno-path Manufacturing is recalling Techno-Path Multichem AE Unassayed Level 3, distributed nationwide in the US. The recall was initiated on 14 Aug 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI:(01)05391523442824/-Lot # 65705263
Quantity recalled832 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0207-2027
FDA event ID
99724
Recalling firm
Techno-path Manufacturing, Ballina Tipperary
Classification
Class II
Status
Ongoing
Recall initiated
14 Aug 2026
FDA classified
22 Oct 2026
Reported
28 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Potency & Assay

Timeline

14 Aug 2026Recall initiated by company
22 Oct 2026Classified by FDA+69 days
28 Oct 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Techno-Path Multichem AE Unassayed Level 3, REF:AE623X

Where it was distributed

US Nationwide distribution in the state of Massachusetts.

Nationwide Massachusetts

Questions about this recall

Why was it recalled?

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

Which lots are affected?

UDI:(01)05391523442824/-Lot # 65705263

Where was it sold?

US Nationwide distribution in the state of Massachusetts.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 3 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.