CLASS II Device recall · Reported 22 Aug 2012 · FDA Enforcement Report

00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner recalled

Zimmer is recalling 00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner 65¿ Neck Angle Per surgical, distributed nationwide in the US. The recall was initiated on 18 Jul 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart 00-4309-028-01: Lot 60529860, 60574377, 60589079, 60605856, 60616574, 60684774, 60842967, 60739623, 60896979, 60936815, 61019785, 61051017
Quantity recalled252 (210 in USA)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2213-2012
FDA event ID
62540
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
18 Jul 2012
FDA classified
16 Aug 2012
Reported
22 Aug 2012
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Foreign Material

Timeline

18 Jul 2012Recall initiated by company
16 Aug 2012Classified by FDA+29 days
22 Aug 2012Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner 40mm 65¿ Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60¿ (Standard) or 65¿ (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."

Where it was distributed

Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.

Nationwide

Questions about this recall

Is 00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner 65¿ Neck Angle Per surgical recalled?

Yes. Zimmer recalled 00-4309-028-01 - Trabecular Metal Reverse Shoulder Liner 65¿ Neck Angle Per surgical on 18 Jul 2012. The recall is Class II and its status is Terminated. Recall number Z-2213-2012.

Why was it recalled?

Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain

Which lots are affected?

Part 00-4309-028-01: Lot 60529860, 60574377, 60589079, 60605856, 60616574, 60684774, 60842967, 60739623, 60896979, 60936815, 61019785, 61051017

Where was it sold?

Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zimmer

All 539

Other Surgical Instruments recalls

All 3,809