CLASS II ONGOING Device recall · Reported 20 Aug 2025 · FDA Enforcement Report

3mensio Workstation (Vascular Fenestrated) recalled

PIE Medical Imaging is recalling 3mensio Workstation (Vascular Fenestrated), distributed nationwide in the US. The recall was initiated on 14 Jul 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware Version/UDI/Software Release Date: 10.6/(01)08056304455505(11)240325(8012)10.6/05-Apr-24, 10.7/(01)08056304455925(11)250423(8012)10.7/29-Apr-25, Service Packs/UDI: 10.6 SP1/(01)08056304455505(11)240411(8012)10.6 SP1/23-Apr-24, 10.6 SP2/(01)08056304455505(11)240610(8012)10.6 SP2/17-Jun-24, 10.6 SP3/(01)08056304455505(11)240926(8012)10.6 SP3/03-Oct-24, 10.6 SP4/(01)08056304455505(11)250109(8012)10.6 SP4/20-Jan-25, 10.7 SP1/(01)08056304455925(11)250523(8012)10.7 SP1/28-May-25
Quantity recalled501

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2297-2025
FDA event ID
97307
Recalling firm
PIE Medical Imaging, Maastricht, N/A
Classification
Class II
Status
Ongoing
Recall initiated
14 Jul 2025
FDA classified
12 Aug 2025
Reported
20 Aug 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

14 Jul 2025Recall initiated by company
12 Aug 2025Classified by FDA+29 days
20 Aug 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

3mensio Workstation (Vascular Fenestrated) software

Where it was distributed

US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.

Nationwide CaliforniaGeorgiaKansasMassachusettsMinnesotaMississippiNorth CarolinaNebraskaNew JerseyNew YorkOhioTexas

Questions about this recall

Why was it recalled?

When fenestrated analysis with clock measurements is started in diagnostic bioimaging software (intended to measure/visualize cardiovascular structures) and the 12h position of a single clock is changed, other clock measurements are not updated relative to the new 12h position which may cause stent graft fenestrations at incorrect position, which may lead to blood flow disruption and tissue damage

Which lots are affected?

Software Version/UDI/Software Release Date: 10.6/(01)08056304455505(11)240325(8012)10.6/05-Apr-24, 10.7/(01)08056304455925(11)250423(8012)10.7/29-Apr-25, Service Packs/UDI: 10.6 SP1/(01)08056304455505(11)240411(8012)10.6 SP1/23-Apr-24, 10.6 SP2/(01)08056304455505(11)240610(8012)10.6 SP2/17-Jun-24, 10.6 SP3/(01)08056304455505(11)240926(8012)10.6 SP3/03-Oct-24, 10.6 SP4/(01)08056304455505(11)250109(8012)10.6 SP4/20-Jan-25, 10.7 SP1/(01)08056304455925(11)250523(8012)10.7 SP1/28-May-25

Where was it sold?

US Nationwide distribution in the states of MN, NJ, GA, NY, OH, KS, MA, CA, TX, MS, NE, NC.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from PIE Medical Imaging

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS II3mensio Workstation (Vascular Fenestrated) recalledPIE Medical ImagingDevice MalfunctionNationwide

Other Heart & Vascular Devices recalls

All 6,199