CLASS II ONGOING Device recall · Reported 28 Aug 2019 · FDA Enforcement Report
Acuson NX3 Elite Diagnostic Ultrasound System recalled
Siemens Medical Solutions USA is recalling Acuson NX3 Elite Diagnostic Ultrasound System, distributed nationwide in the US. The recall was initiated on 28 Jan 2019 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2353-2019
- FDA event ID
- 83398
- Recalling firm
- Siemens Medical Solutions USA, Mountain View, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 28 Jan 2019
- FDA classified
- 22 Aug 2019
- Reported
- 28 Aug 2019
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Heart & Vascular Devices
Timeline
| 28 Jan 2019 | Recall initiated by company | |
| 22 Aug 2019 | Classified by FDA | +206 days |
| 28 Aug 2019 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
ACUSON NX3 Elite Diagnostic Ultrasound System, Model # 11235411, RX Only, Siemens Medical Solutions, containing software versions: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B; UDI # 04056869004303 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Where it was distributed
Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.
Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaIowaIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMississippiNorth Carolina
Questions about this recall
Is Acuson NX3 Elite Diagnostic Ultrasound System recalled?
Yes. Siemens Medical Solutions USA recalled Acuson NX3 Elite Diagnostic Ultrasound System on 28 Jan 2019. The recall is Class II and its status is Ongoing. Recall number Z-2353-2019.
Why was it recalled?
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware
Which lots are affected?
All product lots of the Acuson NX3, model # 11235411 containing the following software version: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B;
Where was it sold?
Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Siemens Medical Solutions USA
All 796| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono biplane recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono floor recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis pheno recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | PA |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |