CLASS II Device recall · Reported 17 Jun 2020 · FDA Enforcement Report
Advia Chemistry Enzymatic Creatinine_2 recalled
Siemens Healthcare Diagnostics is recalling Advia Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, distributed nationwide in the US. The recall was initiated on 4 May 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2282-2020
- FDA event ID
- 85602
- Recalling firm
- Siemens Healthcare Diagnostics, Newark, DE
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 4 May 2020
- FDA classified
- 5 Jun 2020
- Reported
- 17 Jun 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Other Medical Devices
Timeline
| 4 May 2020 | Recall initiated by company | |
| 5 Jun 2020 | Classified by FDA | +32 days |
| 17 Jun 2020 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
ADVIA Chemistry Enzymatic Creatinine_2, Siemens Material Number 10335869, UDI Number 00630414526690 - Product Usage: use in the quantitative determination of creatinine in human serum, plasma, and urine for the diagnosis and treatment of renal disease.
Where it was distributed
Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.
Nationwide AlaskaArkansasArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMinnesota
Questions about this recall
Why was it recalled?
The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.
Which lots are affected?
Lot #s 53149, 53150, 53151, 53152, 53941, 53942, 53943, 53944, 54354, 54371, 54372, 54373, 54374
Where was it sold?
Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
5 other products · FDA event 85602This product is part of a recall event; the main page for the event is Advia Chemistry Fructosamine Reagent (Fruc) recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Advia Chemistry Fructosamine Reagent (Fruc) recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide | |
| CLASS II | Atellica CH Enzymatic Creatinine_2, Siemens Material Number recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide | |
| CLASS II | Atellica CH Fructosamine Reagent (FRUC) recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide | |
| CLASS II | Dimension Enzymatic Creatinine, Siemens Material Number recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide | |
| CLASS II | Dimension Vista Enzymatic Creatinine recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
More recalls from Siemens Healthcare Diagnostics
All 694| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS III | Syva Emit Disopyramide Assay Enzyme Immunoassay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Atellica CH Urine Albumin recalled by Siemens Healthcare Diagnostics | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS III | Syva Emit tox Serum Tricyclic Antidepressant recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |