CLASS II Device recall · Reported 14 Jul 2021 · FDA Enforcement Report

The Advia Chemistry XPT Chemistry System recalled

Siemens Healthcare Diagnostics is recalling The Advia Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an, distributed nationwide in the US. The recall was initiated on 18 May 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware version: 1.4 or lower UDI - 00630414595467
Quantity recalled842 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2014-2021
FDA event ID
88086
Recalling firm
Siemens Healthcare Diagnostics, Tarrytown, NY
Classification
Class II
Status
Terminated
Recall initiated
18 May 2021
FDA classified
6 Jul 2021
Reported
14 Jul 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Spinal Implants

Timeline

18 May 2021Recall initiated by company
6 Jul 2021Classified by FDA+49 days
14 Jul 2021Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes. Tests performed using this system are intended for in vitro diagnostic use.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, MA, MD, MI, MN, NY, OH, PA, TX and the countries of Australia, Bahrain, Brazil, Canada, Chile, Croatia, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Italy, Malaysia, Netherlands, New Zealand, Norway, P.R. China, Portugal, Republic Korea, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.

Nationwide CaliforniaColoradoConnecticutFloridaGeorgiaMassachusettsMarylandMichiganMinnesotaNew YorkOhioPennsylvaniaTexas

Questions about this recall

Why was it recalled?

ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result

Which lots are affected?

Software version: 1.4 or lower UDI - 00630414595467

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, MA, MD, MI, MN, NY, OH, PA, TX and the countries of Australia, Bahrain, Brazil, Canada, Chile, Croatia, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Italy, Malaysia, Netherlands, New Zealand, Norway, P.R. China, Portugal, Republic Korea, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.