CLASS II Device recall · Reported 30 Dec 2015 · FDA Enforcement Report
Aegis Torque Handle recalled by DePuy Spine
DePuy Spine is recalling Aegis Torque Handle, distributed in 9 states. The recall was initiated on 20 Nov 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0454-2016
- FDA event ID
- 72654
- Recalling firm
- DePuy Spine, Raynham, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 20 Nov 2015
- FDA classified
- 21 Dec 2015
- Reported
- 30 Dec 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Processing Deviation
- Device type
- Spinal Implants
Timeline
| 20 Nov 2015 | Recall initiated by company | |
| 21 Dec 2015 | Classified by FDA | +31 days |
| 30 Dec 2015 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length
The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.
Product as listed by the FDA
Aegis Torque Handle CATALOG NO. : 2871-10-800. A spinal surgical instrument used on the AEGIS Anterior Lumbar Plate System.
Where it was distributed
US Distribution to the states of : NC, MI, CO, MD, FL, ID, TX, CA and NY.
CaliforniaColoradoFloridaIdahoMarylandMichiganNorth CarolinaNew YorkTexas
Questions about this recall
Why was it recalled?
Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length
Which lots are affected?
LOT NUMBER: GB0906 GB0707 GB0907 GB58815 GB59244 GB61134
Where was it sold?
US Distribution to the states of : NC, MI, CO, MD, FL, ID, TX, CA and NY.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from DePuy Spine
All 36| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | EIT CIF cage, H, 8degree, L- an intervertebral body fusion recalled | DePuy Spine | Packaging Defects | Nationwide |
| Device | CLASS II | EIT CIF cage, H, 8degree, S- an intervertebral body fusion recalled | DePuy Spine | Packaging Defects | Nationwide |
| Device | CLASS II | Concorde Lift, Lordotic Expandable Interbody Device recalled | DePuy Spine | Other | Nationwide |
| Device | CLASS II | Concorde Lift, Expandable Interbody Device recalled by DePuy Spine | DePuy Spine | Other | Nationwide |
| Device | CLASS II | Confidence Plus Kit Spinal Cement System, Product Codes: 283910000 recalled | DePuy Spine | Other | MA |
Other Spinal Implants recalls
All 1,161| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline surgical convenience kits labeled as: Spinal CDS recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline surgical convenience kits labeled as recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |