CLASS I Device recall · Reported 31 Aug 2016 · FDA Enforcement Report
Alere INRatio2 PT/INR Professional Monitoring System recalled
Alere San Diego is recalling Alere INRatio2 PT/INR Professional Monitoring System, distributed nationwide in the US. The recall was initiated on 11 Jul 2016 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2360-2016
- FDA event ID
- 74665
- Recalling firm
- Alere San Diego, San Diego, CA
- Classification
- Class I
- Status
- Terminated
- Recall initiated
- 11 Jul 2016
- FDA classified
- 19 Aug 2016
- Reported
- 31 Aug 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Press Release
- Reason category
- Other Reasons
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 11 Jul 2016 | Recall initiated by company | |
| 19 Aug 2016 | Classified by FDA | +39 days |
| 31 Aug 2016 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Product as listed by the FDA
Alere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that initiate coagulation. When the blood clots, changes in impedance in the sample are detected by the monitor. The monitor calculates the PT and INR results from this impedance change and reports them on the display. The Alere INRatio/INRatio2 PT/INR Monitoring System is used for the quantitative measurement of Prothrombin Time (PT) in fresh, capillary whole blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for use outside the body (in vitro diagnostic use). The Alere INRatio/INRatio2 PT/INR Monitoring System is intended for professional and home use by people taking warfarin and other oral anticoagulant (blood thinning) therapy who need to monitor the clotting time of their blood. The Alere INRatio/INRatio2 PT/INR Monitoring System is not intended to be used for screening purposes.
Where it was distributed
Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.
Questions about this recall
Is Alere INRatio2 PT/INR Professional Monitoring System recalled?
Yes. Alere San Diego recalled Alere INRatio2 PT/INR Professional Monitoring System on 11 Jul 2016. The recall is Class I and its status is Terminated. Recall number Z-2360-2016.
Why was it recalled?
Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
Which lots are affected?
All Lots All Serial Numbers
Where was it sold?
Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Alere San Diego
All 50| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Abbott Cholestech LDX Battery Kit recalled by Alere San Diego | Alere San Diego | E. coli | Nationwide |
| Device | CLASS II | The Cholestech LDX¿ System combines enzymatic methodology recalled | Alere San Diego | E. coli | 31 states |
| Device | CLASS II | Alere Cholestech LDX Analyzer recalled by Alere San Diego | Alere San Diego | E. coli | Nationwide |
| Device | CLASS III | lmmunalysis Eddp Urine Control Set 1 The Eddp Urine recalled | Alere San Diego | Potency & Assay | 14 states |
| Device | CLASS II | Triage Drugs of Abuse Plus TCA 25 Test Kit recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 459| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Lead & Heavy Metals | Nationwide |
| Device | CLASS I | Omnipod (Eros) Pod recalled by Insulet | Insulet | Lead & Heavy Metals | Nationwide |