CLASS II ONGOING Device recall · Reported 3 Jun 2020 · FDA Enforcement Report
Architect iGentamicin Reagent Kit recalled by Abbott Gmbh & Co. KG
Abbott Gmbh & Co. KG is recalling Architect iGentamicin Reagent Kit, distributed nationwide in the US. The recall was initiated on 24 Apr 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2117-2020
- FDA event ID
- 85590
- Recalling firm
- Abbott Gmbh & Co. KG, Wiesbaden, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 24 Apr 2020
- FDA classified
- 26 May 2020
- Reported
- 3 Jun 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Patient Monitors
Timeline
| 24 Apr 2020 | Recall initiated by company | |
| 26 May 2020 | Classified by FDA | +32 days |
| 3 Jun 2020 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.
Where it was distributed
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CT, DE, FL, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MT, ND, NE, NJ, NV, NY, OH, OK, PA, RI, SC, TN,TX, UT, VA, VT, WA, WI, PR. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Belarus, Belgium, Brazil, Brunei, Canada, Cayman Islands, Chile, China, Costa Rica, Croatia, Curacao, Czech Republic, Faroe Islands, France, Germany, Greece, Greenland, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Namibia, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Philippines, Poland, Portugal, Republic of Yemen, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tunisia, UAE, United Kingdom and Uruguay.
Nationwide AlabamaArizonaCaliforniaConnecticutDelawareFloridaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaineMichiganMinnesotaMissouriMontana
Questions about this recall
Why was it recalled?
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.
Which lots are affected?
MODEL: 1P31-25 LOT NUMBERS: 04281BE00, 08510BE00 and 11221BE00 Shelf life per LOT: 04281BE00;2020-MAY-21 08510BE00; 2020-AUG-08 11221BE00;2020-NOV-18
Where was it sold?
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CT, DE, FL, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MT, ND, NE, NJ, NV, NY, OH, OK, PA, RI, SC, TN,TX, UT, VA, VT, WA, WI, PR. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Belarus, Belgium, Brazil, Brunei, Canada, Cayman Islands, Chile, China, Costa Rica, Croatia, Curacao, Czech Republic, Faroe Islands, France, Germany, Greece, Greenland, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kenya, Latvia, Lebanon, Lithuania, Malaysia, Mauritius, Namibia, Netherlands, New Zealand, Norway, Oman, Palestinian Territory, Philippines, Poland, Portugal, Republic of Yemen, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tunisia, UAE, United Kingdom and Uruguay.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Abbott Gmbh & Co. KG
All 19| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Alinity ci series System Control Module (SCM) recalled | Abbott Gmbh & Co. KG | Device Malfunction | Nationwide |
| Device | CLASS II | Alinity C, Processing Module recalled by Abbott Gmbh & Co. KG | Abbott Gmbh & Co. KG | Potency & Assay | 25 states |
| Device | CLASS II | Architect Havab G Reagent Kit recalled by Abbott Gmbh & Co. KG | Abbott Gmbh & Co. KG | Hepatitis A | Nationwide |
| Device | CLASS II | LN 3R70 recalled by Abbott Gmbh & Co. KG | Abbott Gmbh & Co. KG | Device Malfunction | Nationwide |
| Device | CLASS II | Accessory kit, Alinity c containing Alinity ci Level Sensor recalled | Abbott Gmbh & Co. KG | Device Malfunction | Nationwide |
Other Patient Monitors recalls
All 1,022| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Passport Series Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | T1 Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |