CLASS I Device recall · Reported 8 Jul 2020 · FDA Enforcement Report
Arrow AutoCAT2 Intra-Aortic Balloon Pump Autocat 2, Product recalled
Arrow International is recalling Arrow AutoCAT2 Intra-Aortic Balloon Pump Autocat 2, Product. The recall was initiated on 20 May 2020 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2408-2020
- FDA event ID
- 85684
- Recalling firm
- Arrow International, Reading, PA
- Classification
- Class I
- Status
- Terminated
- Recall initiated
- 20 May 2020
- FDA classified
- 26 Jun 2020
- Reported
- 8 Jul 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Heart & Vascular Devices
Timeline
| 20 May 2020 | Recall initiated by company | |
| 26 Jun 2020 | Classified by FDA | +37 days |
| 8 Jul 2020 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT 2, Product Code IAP-0400
Where it was distributed
Worldwide distribution.
Questions about this recall
Is Arrow AutoCAT2 Intra-Aortic Balloon Pump Autocat 2, Product recalled?
Yes. Arrow International recalled Arrow AutoCAT2 Intra-Aortic Balloon Pump Autocat 2, Product on 20 May 2020. The recall is Class I and its status is Terminated. Recall number Z-2408-2020.
Why was it recalled?
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Which lots are affected?
Serial number ranges: Arrow AutoCAT 2 Intra-Aortic Balloon Pump AUTOCAT 2 IAP-0400 140868V 140901V through 140909V 140914V through 140920V 150102V 150106V through 150107V 150110V through 150113V 150115V through 150116V 150220V 150401V 150407V 150409V 150507V through 150513V 150601V through 150605V 150608V through 150611V 150625V 150702V through 150705V 150707V 150720V 150801V through 150802V 150821V 150824V 150827V through 150830V 150901V through 150906V 150908V through 150912V 150919V through 150923V 150925V through 150939V 151009V through 151013V 151015V through 151016V 151019V 151021V through 151027V 151029V 151102V through 151103V 151105V 151108V 151111V 151113V 151201V through 151208V 151210V through 151212V 151214V through 151218V Arrow AutoCAT 2 Intra-Aortic Balloon Pump AUTOCAT 2 IAP-0400 151220V 160101V through 160103V 160105V through 160109V 160112V through 160118V 160120V 160122V 160126V 160201V through 160209V 160211V 160213V through 160214V 160216V through 160217V 160219V 160312V through 160314V 160316V through 160323V 160325V through 160327V 160331V 160401V through 160410V 160412V 160418V through 160422V 160519V 160525V 160529V through 160530V 160605V 160614V 160618V 160630V 160638V through 160639V 160645V 160703V 160718V 160720V 160726V 160728V through 160730V 170101V through 170110V 170401V through 170410V 170501V through 170520V 170601V through 170625V 170811V through 170830V 171021V through 171030V 171037V through 171046V 171106V through 171115V 180131V through 180150V Arrow AutoCAT 2 Intra-Aortic Balloon Pump AUTOCAT 2 IAP-0400 180211V through 180230V 180411V through 180420V 180441V through 180450V 180521V through 180530V 180621V through 180630V 180711V through 180720V 180741V through 180750V 180811V through 180820V 181011V through 181040V 181111V through 181130V 190121V through 190130V 190141V through 190150V 190211V through 190220V 190411V through 190420V 190431V through 190440V 190531V through 190540V 190701V through 190712V 190737V through 190748V 190813V through 190824V 190837V through 190848V 190913V through 190924V 190937V through 190948V 191001V through 191012V 191025V through 191036V 191101V through 191112V 191137V through 191148V 200113V through 200136V 200314V through 200324V
Where was it sold?
Worldwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Arrow International
All 525| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Description/REF: Agba Picc G4 Stylet: 3L 6FR x 55CM recalled | Arrow International | Lead & Heavy Metals | Nationwide |
| Device | CLASS I | Description/REF: Hemodialysis KIT: 2-Lumen 8 FR X 11 CM/AK-11802-F recalled | Arrow International | Lead & Heavy Metals | Nationwide |
| Device | CLASS I | Description/REF: PI CVC KIT recalled by Arrow International | Arrow International | Lead & Heavy Metals | Nationwide |
| Device | CLASS I | Description/REF: PI Picc: 3: 6FR x 55CM w CHG/ Dlx-35563-Hpkc; Agba recalled | Arrow International | Lead & Heavy Metals | Nationwide |
| Device | CLASS I | Description/REF: Pacing PSI KIT: 6 FR/AI-06210-KS, Pacing PSI KIT recalled | Arrow International | Lead & Heavy Metals | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |