CLASS II Device recall · Reported 11 Nov 2015 · FDA Enforcement Report

Arrow PICC Kit featuring Chlorag+ard Technology Midline recalled

Arrow International is recalling Arrow PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit Arrow, distributed nationwide in the US. The recall was initiated on 16 Oct 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Codes ASK-45552-RWJ3 CDC-35541-HPK1A EU-00750-45 PL-00750 AU-00420-MAJO CDC-35541-VPS EU-02041-ML PL-00755 CDC-34041-HPK1A CDC-35552-HPK1A EU-24041-HPMSB PR-35552-HPHNM CDC-34041-VPS CDC-35552-VPS EU-24052-HPMSB PR-45563-HPHNM CDC-34052-HPK1A CDC-41563-JX1A EU-25041-HPMSB UK-00420-MID CDC-34052-VPS CDC-44041-HPK1A EU-25052-HPMSB UK-00420-RCHT CDC-35041-HPK1A CDC-44052-HPK1A EU-25063-HPMSB UK-00420-THT CDC-35041-VPS CDC-44063-HPK1A EU-25541-HPMSB UK-04041-RDEH CDC-35052-HPK1A CDC-45063-HPK1A EU-25552-HPMSB UK-05041-MSB CDC-35052-VPS CDC-45563-HPK1A IB-01652 UK-05041-RDEH CDC-35063-HPK1A EU-00740-45 PL-00740 UK-05052-RCHT
Quantity recalled12,157

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0207-2016
FDA event ID
72463
Recalling firm
Arrow International, Reading, PA
Classification
Class II
Status
Terminated
Recall initiated
16 Oct 2015
FDA classified
4 Nov 2015
Reported
11 Nov 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

16 Oct 2015Recall initiated by company
4 Nov 2015Classified by FDA+19 days
11 Nov 2015Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter. Although the issue does not pose a significant risk to the patient, it may cause a delay of treatment when placing the catheter.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ARROW PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit ARROW Advantage5 Pressure Injectable PICC Kit Arrow¿ Pressure Injectable Arrow¿ PICC powered by Arrow¿ VPS¿ Stylet ARROW Pressure Injectable JACC with Chlorag+ard Technology For short-term or long-term peripheral access to the central venous system for intravenous therapy, blood sampling, infusion, pressure injection of contrast media, and allows for central venous pressure monitoring. Permits venous access to the peripheral circulation. Offers an alternative method of intravenous access for select adult and pediatric patients. The ARROW GlideThru Peel-Away Sheath/Dilator is used for the percutaneous introduction of diagnostic or therapeutic devices into the vasculature.

Where it was distributed

World Wide Distribution - US including AR, AL, TX, MA, CA, IA, MO, NY, AL, VA, RI, IN, TN, FL, KY, NH, SC, MD, WA, OK, IL, ME, OR, OH, PA, MI, KS, LA, TN, WV, NE, NJ, NV, GA, SD, CT, MS, CO, and Internationally to Canada and Belgium.

Nationwide AlabamaArkansasCaliforniaColoradoConnecticutFloridaGeorgiaIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMissouriMississippiNebraska

Questions about this recall

Is Arrow PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit Arrow recalled?

Yes. Arrow International recalled Arrow PICC Kit featuring Chlorag+ard Technology Midline Catheterization Kit Arrow on 16 Oct 2015. The recall is Class II and its status is Terminated. Recall number Z-0207-2016.

Why was it recalled?

Possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel apart the sheath body from the catheter. Although the issue does not pose a significant risk to the patient, it may cause a delay of treatment when placing the catheter.

Which lots are affected?

Product Codes ASK-45552-RWJ3 CDC-35541-HPK1A EU-00750-45 PL-00750 AU-00420-MAJO CDC-35541-VPS EU-02041-ML PL-00755 CDC-34041-HPK1A CDC-35552-HPK1A EU-24041-HPMSB PR-35552-HPHNM CDC-34041-VPS CDC-35552-VPS EU-24052-HPMSB PR-45563-HPHNM CDC-34052-HPK1A CDC-41563-JX1A EU-25041-HPMSB UK-00420-MID CDC-34052-VPS CDC-44041-HPK1A EU-25052-HPMSB UK-00420-RCHT CDC-35041-HPK1A CDC-44052-HPK1A EU-25063-HPMSB UK-00420-THT CDC-35041-VPS CDC-44063-HPK1A EU-25541-HPMSB UK-04041-RDEH CDC-35052-HPK1A CDC-45063-HPK1A EU-25552-HPMSB UK-05041-MSB CDC-35052-VPS CDC-45563-HPK1A IB-01652 UK-05041-RDEH CDC-35063-HPK1A EU-00740-45 PL-00740 UK-05052-RCHT

Where was it sold?

World Wide Distribution - US including AR, AL, TX, MA, CA, IA, MO, NY, AL, VA, RI, IN, TN, FL, KY, NH, SC, MD, WA, OK, IL, ME, OR, OH, PA, MI, KS, LA, TN, WV, NE, NJ, NV, GA, SD, CT, MS, CO, and Internationally to Canada and Belgium.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Arrow International

All 525

Other Heart & Vascular Devices recalls

All 6,190