CLASS II ONGOING Device recall · Reported 27 Apr 2022 · FDA Enforcement Report

Artis Q.zen floor recalled by Siemens Medical Solutions USA

Siemens Medical Solutions USA is recalling Artis Q.zen floor, distributed nationwide in the US. The recall was initiated on 23 Feb 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 4056869010014; Serial Numbers: 105120 105110 105033 105032 105500 105105 105028 105503 105018 105114 105115 105026 105025 105135 105016 105119 105137 105501 105132 105128 105142 105017 105121 105122 105123 105100 105112 105124 105029 105030 105031 105125 Additional Serial Numbers as of 10/4/23: 143500
Quantity recalled33 systems

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0978-2022
FDA event ID
89860
Recalling firm
Siemens Medical Solutions USA, Malvern, PA
Classification
Class II
Status
Ongoing
Recall initiated
23 Feb 2022
FDA classified
21 Apr 2022
Reported
27 Apr 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

23 Feb 2022Recall initiated by company
21 Apr 2022Classified by FDA+57 days
27 Apr 2022Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Artis Q.zen floor, Model Number 10848353

Where it was distributed

US Nationwide.

Nationwide

Questions about this recall

Why was it recalled?

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Which lots are affected?

UDI: 4056869010014; Serial Numbers: 105120 105110 105033 105032 105500 105105 105028 105503 105018 105114 105115 105026 105025 105135 105016 105119 105137 105501 105132 105128 105142 105017 105121 105122 105123 105100 105112 105124 105029 105030 105031 105125 Additional Serial Numbers as of 10/4/23: 143500

Where was it sold?

US Nationwide.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

13 other products · FDA event 89860

This product is part of a recall event; the main page for the event is Artis zee floor MN recalled by Siemens Medical Solutions USA.

ReportedClassRecallCompanyReasonWhere
CLASS IIArtis zee floor MN recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis Q ceiling recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis Q zeego recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis zee biplane recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis zee multi-purpose recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis zee biplane MN recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis Q floor recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis zee ceiling recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis Q.zen biplane recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis Q biplane recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis Q.zen ceiling recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis zee floor recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide
CLASS IIArtis zeego recalled by Siemens Medical Solutions USASiemens Medical Solutions USADevice MalfunctionNationwide

More recalls from Siemens Medical Solutions USA

All 796

Other Other Medical Devices recalls

All 13,778