CLASS II Device recall · Reported 2 Aug 2017 · FDA Enforcement Report

Atrium iCAST Balloon Expandable Covered Stent recalled by Atrium Medical

Atrium Medical is recalling Atrium iCAST Balloon Expandable Covered Stent, distributed nationwide in the US. The recall was initiated on 25 May 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Number 85455: Lots 240735, 240523; Part Number 85355: Lot 240734
Quantity recalled246 (US: 183 units; Foreign: 63 units)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2761-2017
FDA event ID
77342
Recalling firm
Atrium Medical, Hudson, NH
Classification
Class II
Status
Terminated
Recall initiated
25 May 2017
FDA classified
25 Jul 2017
Reported
2 Aug 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

25 May 2017Recall initiated by company
25 Jul 2017Classified by FDA+61 days
2 Aug 2017Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm received a complaint that a customer who ordered an ADVANTA V12 7mm x 22mm x 120cm received an ADVANTA v12 7 mm x 38 mm x 80 cm, due to the manifold (catheter hub) was printed as 7mm x 38 mm x 80 cm. The complaint was the result of incorrect dimensional information (7mm x 38mm x 80cm) printed on the manifiold (catheter hub). The complaint device physical dimensions matched the dimensional information displayed on the outer packaging and the device inner pouch. This was the only complaint received for this non-conformance.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85355 (Marketed OUTSIDE U.S. only) The iCAST Covered Stent is indicated for the treatment of tracheobronchial strictures produced by malignant neoplasms.

Where it was distributed

Worldwide distribution-US (nationwide) and countries of: Germany, Great Britain, Austria, Ireland, Netherlands, France, Italy, Sweden, and San Marino.

Nationwide

Questions about this recall

Why was it recalled?

Firm received a complaint that a customer who ordered an ADVANTA V12 7mm x 22mm x 120cm received an ADVANTA v12 7 mm x 38 mm x 80 cm, due to the manifold (catheter hub) was printed as 7mm x 38 mm x 80 cm. The complaint was the result of incorrect dimensional information (7mm x 38mm x 80cm) printed on the manifiold (catheter hub). The complaint device physical dimensions matched the dimensional information displayed on the outer packaging and the device inner pouch. This was the only complaint received for this non-conformance.

Which lots are affected?

Part Number 85455: Lots 240735, 240523; Part Number 85355: Lot 240734

Where was it sold?

Worldwide distribution-US (nationwide) and countries of: Germany, Great Britain, Austria, Ireland, Netherlands, France, Italy, Sweden, and San Marino.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Atrium Medical

All 153

Other Heart & Vascular Devices recalls

All 6,199