CLASS I Device recall · Reported 15 Aug 2012 · FDA Enforcement Report
Automix 3+3 Compounder System, product 2M8286 (115V) recalled
Baxter Healthcare is recalling Automix 3+3 Compounder System, product 2M8286 (115V) and 2M8286K (230V), automated, distributed nationwide in the US. The recall was initiated on 13 Jul 2012 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2141-2012
- FDA event ID
- 62558
- Recalling firm
- Baxter Healthcare, Round Lake, IL
- Classification
- Class I
- Status
- Terminated
- Recall initiated
- 13 Jul 2012
- FDA classified
- 8 Aug 2012
- Reported
- 15 Aug 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 13 Jul 2012 | Recall initiated by company | |
| 8 Aug 2012 | Classified by FDA | +26 days |
| 15 Aug 2012 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
AUTOMIX 3+3 Compounder System, product 2M8286 (115V) and 2M8286K (230V), automated nutrition compounders; Product Usage: The Automix Compounder Systems are electromechanical positive displacement fluid compounding devices used in the admixture of large volume parenteral nutrition formulations via gravimetric weighing.
Where it was distributed
Worldwide Distribution - USA - Nationwide including Puerto Rico, and internationally to Brazil, Canada, Colombia, Costa Rica, Czech Republic, El Salvador, Egypt, Germany, Guatemala, Honduras, Luxembourg, Mexico, Panama, Saudi Arabia, Singapore, Switzerland, Taiwan, Trinidad, Turkey, United Arab Emirates and United Kingdom.
Questions about this recall
Why was it recalled?
After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
Which lots are affected?
Product codes 2M8286 and 2M8286K, all serial numbers
Where was it sold?
Worldwide Distribution - USA - Nationwide including Puerto Rico, and internationally to Brazil, Canada, Colombia, Costa Rica, Czech Republic, El Salvador, Egypt, Germany, Guatemala, Honduras, Luxembourg, Mexico, Panama, Saudi Arabia, Singapore, Switzerland, Taiwan, Trinidad, Turkey, United Arab Emirates and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
3 other products · FDA event 62558This product is part of a recall event; the main page for the event is Automix High Speed Compounder System, product 2M8077 recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS I | Automix High Speed Compounder System, product 2M8077 recalled | Baxter Healthcare | Device Malfunction | Nationwide | |
| CLASS I | Automix 3+3 Compounder with Accusource Monitoring System recalled | Baxter Healthcare | Device Malfunction | Nationwide | |
| CLASS I | Automix Plus High Speed Compounder System, product 2M8075 recalled | Baxter Healthcare | Device Malfunction | Nationwide |
More recalls from Baxter Healthcare
All 573| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Drug | CLASS I | Dextrose Injection, USP, 70 % recalled by Baxter Healthcare | Baxter Healthcare | Particulate Matter | Nationwide |
| Drug | CLASS I | 0.9% Sodium Chloride Injection USP, Viaflex recalledSodium Chloride | Baxter Healthcare | Particulate Matter | IL |
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Drug | CLASS II | Clindamycin Phosphate in 0.9% Sodium Chloride Injection recalledClindamycin Phosphate | Baxter Healthcare | Manufacturing (CGMP) Deviations | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |