CLASS II ONGOING Device recall · Reported 21 Dec 2022 · FDA Enforcement Report
BBL Chocolate II Agar- IVD for the isolation recalled
Becton Dickinson & is recalling BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species, distributed nationwide in the US. The recall was initiated on 21 Nov 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0473-2023
- FDA event ID
- 91201
- Recalling firm
- Becton Dickinson &, Sparks, MD
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 21 Nov 2022
- FDA classified
- 9 Dec 2022
- Reported
- 21 Dec 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 21 Nov 2022 | Recall initiated by company | |
| 9 Dec 2022 | Classified by FDA | +18 days |
| 21 Dec 2022 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Canada, India, Philippines.
Questions about this recall
Is BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species recalled?
Yes. Becton Dickinson & recalled BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species on 21 Nov 2022. The recall is Class II and its status is Ongoing. Recall number Z-0473-2023.
Why was it recalled?
Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.
Which lots are affected?
UDI: (10)2206461(17)221125(30)0100 Lot Number: 2206461 Exp. Date: 20221125. Box numbers 0002 through 0097. Boxes correlate to timestamps on the plates of 11:34 through 13:51. All other box numbers and plate stamps are unaffected and perform as intended. (Timestamps can be found on the bottom of the plate. The box number can be located on the carton label).
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Canada, India, Philippines.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Becton Dickinson &
All 361| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle recalled by Becton Dickinson & | Becton Dickinson & | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle recalled by Becton Dickinson & | Becton Dickinson & | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Quincke Needle recalled by Becton Dickinson & | Becton Dickinson & | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle recalled by Becton Dickinson & | Becton Dickinson & | Processing Deviation | Nationwide |
| Device | CLASS I | BD¿Spinal Tray with BD¿Whitacre Needle, BD¿Quincke Needle recalled | Becton Dickinson & | Processing Deviation | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |