CLASS II ONGOING Device recall · Reported 13 Dec 2023 · FDA Enforcement Report
Biofire Blood Culture Identification 2 (BCID2) Panel recalled
BioFire Diagnostics is recalling Biofire Blood Culture Identification 2 (BCID2) Panel, distributed nationwide in the US. The recall was initiated on 17 Oct 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0451-2024
- FDA event ID
- 93360
- Recalling firm
- BioFire Diagnostics, Salt Lake City, UT
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 17 Oct 2023
- FDA classified
- 1 Dec 2023
- Reported
- 13 Dec 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Other Medical Devices
Timeline
| 17 Oct 2023 | Recall initiated by company | |
| 1 Dec 2023 | Classified by FDA | +45 days |
| 13 Dec 2023 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251
Where it was distributed
Worldwide Distribution: US (nationwide): DE, OH, NY, IL, KY, FL, GA, KS, CA, WI, NC,, AK, PA, MN, ND, IN, AR, CO, UT, TX, AL, NJ, MD, SD, LA, OR, MA, ID, NE, MT, VA, IA, MO, WV, TN, MI, WY, SC, AZ, WA, NH, OK, HI, MS, DC, VT, CT, NM, RI, ME, NV; and OUS (foreign) to countries of: Algeria, Angola, Armenia, Bahrain, Bolivia, Bosnia- Herzegovina, Brunei, Bulgaria, Costa Rica, Croatia, Cyprus, Dominican Republic, Dutch Antilles, Ecuador, El Salvador, Estonia, Gabon, Georgia, Guam, Guatemala, Honduras, Indonesia, Ireland, Iceland, Israel, Ivory Coast, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macao, Macedonia, Malaysia, Mali, Mongolia, Morocco, New Zealand, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Qatar, Romania, Saudi Arabia, Serbia , Slovenia, Tunisia, Uganda, United Arab Emirate, Uruguay, Venezuela, Vietnam, Zimbabwe, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Spain, France, United Kingdom, Greece, India, Italia, Japan, Korea, Netherlands, Poland, Portugal, Sweden, Singapore, Thailand, Turkey, Taiwan, Switzerland, Austria, Denmark, Finland, Germany, Belgium, Hungary, South Africa, Philippines, Norway, Mexico
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.
Which lots are affected?
REF/UDI-DI: RFIT-ASY-0147/00815381020338. FLM1-PRT-0251-16 and prior.
Where was it sold?
Worldwide Distribution: US (nationwide): DE, OH, NY, IL, KY, FL, GA, KS, CA, WI, NC,, AK, PA, MN, ND, IN, AR, CO, UT, TX, AL, NJ, MD, SD, LA, OR, MA, ID, NE, MT, VA, IA, MO, WV, TN, MI, WY, SC, AZ, WA, NH, OK, HI, MS, DC, VT, CT, NM, RI, ME, NV; and OUS (foreign) to countries of: Algeria, Angola, Armenia, Bahrain, Bolivia, Bosnia- Herzegovina, Brunei, Bulgaria, Costa Rica, Croatia, Cyprus, Dominican Republic, Dutch Antilles, Ecuador, El Salvador, Estonia, Gabon, Georgia, Guam, Guatemala, Honduras, Indonesia, Ireland, Iceland, Israel, Ivory Coast, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macao, Macedonia, Malaysia, Mali, Mongolia, Morocco, New Zealand, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Qatar, Romania, Saudi Arabia, Serbia , Slovenia, Tunisia, Uganda, United Arab Emirate, Uruguay, Venezuela, Vietnam, Zimbabwe, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Spain, France, United Kingdom, Greece, India, Italia, Japan, Korea, Netherlands, Poland, Portugal, Sweden, Singapore, Thailand, Turkey, Taiwan, Switzerland, Austria, Denmark, Finland, Germany, Belgium, Hungary, South Africa, Philippines, Norway, Mexico
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from BioFire Diagnostics
All 47| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Multiple-type respiratory pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Multiple-type respiratory pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Mulitple-type osteoarticular pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 2.1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |
| Device | CLASS II | Biofire Spotfire Respiratory/Sore Throat Panel recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,929| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Rapicide PA Part A High-Level Disinfectant recalled by Steris | Steris | Other | Nationwide |
| Device | CLASS II | NeuroStar Advanced Therapy System recalled by Neuronetics | Neuronetics | Other | Nationwide |
| Device | CLASS II | TS Misc. Instruments Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |