CLASS II Device recall · Reported 30 Mar 2022 · FDA Enforcement Report

Biomet Trauma, [Item # 13182-22-26, DVR Crosslock Extra Long Plate recalled

Biomet is recalling Biomet Trauma, [, distributed in 28 states. The recall was initiated on 14 Feb 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes6 lots: Lot #, UDI - [101220, (01)00887868446191(17)310623(10)101220], [101340, (01)00887868446207(17)310625(10)101340], [101450, (01)00887868446313(17)310630(10)101450], [101560, (01)00887868446191(17)310624(10)101560], [119180, (01)00887868446313(17)310630(10)119180], [154060, (01)00887868446313(17)310630(10)154060]
Quantity recalled38

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0810-2022
FDA event ID
89701
Recalling firm
Biomet, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
14 Feb 2022
FDA classified
18 Mar 2022
Reported
30 Mar 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

14 Feb 2022Recall initiated by company
18 Mar 2022Classified by FDA+32 days
30 Mar 2022Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

BIOMET Trauma, [Item # 13182-22-26, DVR CROSSLOCK EXTRA LONG PLATE LEFT], [Item # 1318-12-126, DVR CROSSLOCK EXTRA LONG PLATE RIGHT], [Item # 1318-12-176, DVR CROSSLOCK EXTRA EXTRA LONG PLATE RIGHT]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.

Where it was distributed

Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands

AlabamaArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIdahoIllinoisIndianaKansasKentuckyLouisianaMarylandMichiganMississippiNorth CarolinaNebraskaNew MexicoNew York

Questions about this recall

Is Biomet Trauma, [Item # 13182-22-26, DVR Crosslock Extra Long Plate recalled?

Yes. Biomet recalled Biomet Trauma, [Item # 13182-22-26, DVR Crosslock Extra Long Plate on 14 Feb 2022. The recall is Class II and its status is Terminated. Recall number Z-0810-2022.

Why was it recalled?

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

Which lots are affected?

6 lots: Lot #, UDI - [101220, (01)00887868446191(17)310623(10)101220], [101340, (01)00887868446207(17)310625(10)101340], [101450, (01)00887868446313(17)310630(10)101450], [101560, (01)00887868446191(17)310624(10)101560], [119180, (01)00887868446313(17)310630(10)119180], [154060, (01)00887868446313(17)310630(10)154060]

Where was it sold?

Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Biomet

All 279

Other Other Medical Devices recalls

All 13,778