CLASS II ONGOING Device recall · Reported 17 Jan 2024 · FDA Enforcement Report

Boston Scientific Model 3120 Zoom Latitude recalled

Boston Scientific is recalling Boston Scientific, distributed nationwide in the US. The recall was initiated on 30 Nov 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll serial numbers. GTIN numbers in the U.S.: 00802526434723, 00802526456961, 00802526467653, 00802526468582, 00802526473449, 00802526474859, 00802526475788, 00802526477676, 00802526478239, 00802526479663, 00802526480638, 00802526484599, 00802526486906, 00802526494406, 00802526497605, 00802526516009, 00802526523212, 00802526535406, 00802526554209, 00802526562600, 00802526564703, 00802526575303, 00802526579707, 00802526582004, 00802526602900, 00802526604584, 00802526606366, 00802526606403, and 00802526611322. GTIN numbers OUS: 00802526435430, 00802526468582, 00802526479663, 00802526480638, 00802526486906, 00802526478246, 00802526479670, 00802526484605, 00802526486913, 00802526496608, 00802526496615, 00802526506703, 00802526515903, 00802526517112, 00802526526602, 00802526527609, 00802526554803, 00802526561900, 00802526564802, 00802526574801, 00802526579905, 00802526582011, 00802526602917, 00802526605406, 00802526606342, 00802526611308, 00802526615603, 00802526437458, 00802526441974, 00802526444814, 00802526467288, 00802526469343, 00802526475450, 00802526475757, 00802526477669, 00802526478253, 00802526479687, 00802526484612, 00802526486920, 00802526496622, 00802526496639, 00802526515910, 00802526517136, 00802526526800, 00802526527616, 00802526554810, 00802526564819, 00802526574818, 00802526581021, 00802526582028, 00802526605437, 00802526606311, 00802526612428, 00802526615634, 00802526454301, 00802526467295, 00802526469350, 00802526475467, 00802526475764, 00802526477676, 00802526484803, 00802526486081, 00802526486937, 00802526497629, 00802526515927, 00802526532832, 00802526557002, 00802526565427, 00802526582301, 00802526587405, 00802526602924, 00802526606359, 00802526612442, 00802526445262, 00802526454950, 00802526474170, 00802526475474, 00802526475771, 00802526480652, 00802526486944, 00802526497643, 00802526517174, 00802526526817, 00802526527623, 00802526557019, 00802526565304, 00802526579929, 00802526583803, 00802526602931, 00802526605468, 00802526606304, 00802526612411, 00802526615627, 00802526442827, 00802526444807, 00802526469374, 00802526477683, 00802526482267, 00802526497650, 00802526517198, 00802526540400, 00802526555169, 00802526564826, 00802526582325, 00802526585104, 00802526586316, 00802526586316, 00802526603013, 00802526606328, 00802526612435, 00802526615429, 00802526474057, 00802526477522, 00802526479649, 00802526482274, 00802526486470, 00802526497667, 00802526517211, 00802526540509, 00802526555206, 00802526564833, 00802526582318, 00802526585111, 00802526586408, 00802526603006, 00802526605420, 00802526606335, 00802526612404, 00802526615504, 00802526480669, 00802526605451, 00802526496653, 00802526526824, 00802526527630, 00802526527630, 00802526564840, 00802526581038, 00802526582035, 00802526584602, 00802526582035, 00802526605451, 00802526606403, and 00802526612459.
Quantity recalled14,328 ZOOM programmers

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0673-2024
FDA event ID
93519
Recalling firm
Boston Scientific, Saint Paul, MN
Classification
Class II
Status
Ongoing
Recall initiated
30 Nov 2023
FDA classified
5 Jan 2024
Reported
17 Jan 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 Nov 2023Recall initiated by company
5 Jan 2024Classified by FDA+36 days
17 Jan 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus.

Where it was distributed

Distribution was nationwide, including Guam, Puerto Rico, and the U.S. Virgin Islands. There was also government/military distribution. Foreign distribution was made to Andorra, Argentina, Aruba, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, British Virgin Islands, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Faroe Islands, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Yemen, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan R.O.C., Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, Uruguay, Venezuela, Vietnam, and Zimbabwe.

Nationwide Puerto Rico

Questions about this recall

Is Boston Scientific Model 3120 Zoom Latitude recalled?

Yes. Boston Scientific recalled Boston Scientific Model 3120 Zoom Latitude on 30 Nov 2023. The recall is Class II and its status is Ongoing. Recall number Z-0673-2024.

Why was it recalled?

The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.

Which lots are affected?

All serial numbers. GTIN numbers in the U.S.: 00802526434723, 00802526456961, 00802526467653, 00802526468582, 00802526473449, 00802526474859, 00802526475788, 00802526477676, 00802526478239, 00802526479663, 00802526480638, 00802526484599, 00802526486906, 00802526494406, 00802526497605, 00802526516009, 00802526523212, 00802526535406, 00802526554209, 00802526562600, 00802526564703, 00802526575303, 00802526579707, 00802526582004, 00802526602900, 00802526604584, 00802526606366, 00802526606403, and 00802526611322. GTIN numbers OUS: 00802526435430, 00802526468582, 00802526479663, 00802526480638, 00802526486906, 00802526478246, 00802526479670, 00802526484605, 00802526486913, 00802526496608, 00802526496615, 00802526506703, 00802526515903, 00802526517112, 00802526526602, 00802526527609, 00802526554803, 00802526561900, 00802526564802, 00802526574801, 00802526579905, 00802526582011, 00802526602917, 00802526605406, 00802526606342, 00802526611308, 00802526615603, 00802526437458, 00802526441974, 00802526444814, 00802526467288, 00802526469343, 00802526475450, 00802526475757, 00802526477669, 00802526478253, 00802526479687, 00802526484612, 00802526486920, 00802526496622, 00802526496639, 00802526515910, 00802526517136, 00802526526800, 00802526527616, 00802526554810, 00802526564819, 00802526574818, 00802526581021, 00802526582028, 00802526605437, 00802526606311, 00802526612428, 00802526615634, 00802526454301, 00802526467295, 00802526469350, 00802526475467, 00802526475764, 00802526477676, 00802526484803, 00802526486081, 00802526486937, 00802526497629, 00802526515927, 00802526532832, 00802526557002, 00802526565427, 00802526582301, 00802526587405, 00802526602924, 00802526606359, 00802526612442, 00802526445262, 00802526454950, 00802526474170, 00802526475474, 00802526475771, 00802526480652, 00802526486944, 00802526497643, 00802526517174, 00802526526817, 00802526527623, 00802526557019, 00802526565304, 00802526579929, 00802526583803, 00802526602931, 00802526605468, 00802526606304, 00802526612411, 00802526615627, 00802526442827, 00802526444807, 00802526469374, 00802526477683, 00802526482267, 00802526497650, 00802526517198, 00802526540400, 00802526555169, 00802526564826, 00802526582325, 00802526585104, 00802526586316, 00802526586316, 00802526603013, 00802526606328, 00802526612435, 00802526615429, 00802526474057, 00802526477522, 00802526479649, 00802526482274, 00802526486470, 00802526497667, 00802526517211, 00802526540509, 00802526555206, 00802526564833, 00802526582318, 00802526585111, 00802526586408, 00802526603006, 00802526605420, 00802526606335, 00802526612404, 00802526615504, 00802526480669, 00802526605451, 00802526496653, 00802526526824, 00802526527630, 00802526527630, 00802526564840, 00802526581038, 00802526582035, 00802526584602, 00802526582035, 00802526605451, 00802526606403, and 00802526612459.

Where was it sold?

Distribution was nationwide, including Guam, Puerto Rico, and the U.S. Virgin Islands. There was also government/military distribution. Foreign distribution was made to Andorra, Argentina, Aruba, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, British Virgin Islands, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Faroe Islands, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Yemen, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan R.O.C., Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, Uruguay, Venezuela, Vietnam, and Zimbabwe.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Boston Scientific

All 626

Other Pacemakers & Defibrillators recalls

All 494