CLASS II ONGOING Device recall · Reported 17 Jan 2024 · FDA Enforcement Report
Boston Scientific Model 3120 Zoom Latitude recalled
Boston Scientific is recalling Boston Scientific, distributed nationwide in the US. The recall was initiated on 30 Nov 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0673-2024
- FDA event ID
- 93519
- Recalling firm
- Boston Scientific, Saint Paul, MN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 30 Nov 2023
- FDA classified
- 5 Jan 2024
- Reported
- 17 Jan 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Pacemakers & Defibrillators
Timeline
| 30 Nov 2023 | Recall initiated by company | |
| 5 Jan 2024 | Classified by FDA | +36 days |
| 17 Jan 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus. ALTRUA 20, ALTRUA 40, ALTRUA 60, INSIGNIA I Entra, INSIGNIA I Plus, INSIGNIA I Ultra, NEXUS I Entra, and NEXUS I Plus.
Where it was distributed
Distribution was nationwide, including Guam, Puerto Rico, and the U.S. Virgin Islands. There was also government/military distribution. Foreign distribution was made to Andorra, Argentina, Aruba, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, British Virgin Islands, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Faroe Islands, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Yemen, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan R.O.C., Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, Uruguay, Venezuela, Vietnam, and Zimbabwe.
Questions about this recall
Is Boston Scientific Model 3120 Zoom Latitude recalled?
Yes. Boston Scientific recalled Boston Scientific Model 3120 Zoom Latitude on 30 Nov 2023. The recall is Class II and its status is Ongoing. Recall number Z-0673-2024.
Why was it recalled?
The software application on the Programmer/Recorder/Monitor (PRM) for Implantable Pulse Generators will display certain diagnostic dates incorrectly, with a year in the 1990s.
Which lots are affected?
All serial numbers. GTIN numbers in the U.S.: 00802526434723, 00802526456961, 00802526467653, 00802526468582, 00802526473449, 00802526474859, 00802526475788, 00802526477676, 00802526478239, 00802526479663, 00802526480638, 00802526484599, 00802526486906, 00802526494406, 00802526497605, 00802526516009, 00802526523212, 00802526535406, 00802526554209, 00802526562600, 00802526564703, 00802526575303, 00802526579707, 00802526582004, 00802526602900, 00802526604584, 00802526606366, 00802526606403, and 00802526611322. GTIN numbers OUS: 00802526435430, 00802526468582, 00802526479663, 00802526480638, 00802526486906, 00802526478246, 00802526479670, 00802526484605, 00802526486913, 00802526496608, 00802526496615, 00802526506703, 00802526515903, 00802526517112, 00802526526602, 00802526527609, 00802526554803, 00802526561900, 00802526564802, 00802526574801, 00802526579905, 00802526582011, 00802526602917, 00802526605406, 00802526606342, 00802526611308, 00802526615603, 00802526437458, 00802526441974, 00802526444814, 00802526467288, 00802526469343, 00802526475450, 00802526475757, 00802526477669, 00802526478253, 00802526479687, 00802526484612, 00802526486920, 00802526496622, 00802526496639, 00802526515910, 00802526517136, 00802526526800, 00802526527616, 00802526554810, 00802526564819, 00802526574818, 00802526581021, 00802526582028, 00802526605437, 00802526606311, 00802526612428, 00802526615634, 00802526454301, 00802526467295, 00802526469350, 00802526475467, 00802526475764, 00802526477676, 00802526484803, 00802526486081, 00802526486937, 00802526497629, 00802526515927, 00802526532832, 00802526557002, 00802526565427, 00802526582301, 00802526587405, 00802526602924, 00802526606359, 00802526612442, 00802526445262, 00802526454950, 00802526474170, 00802526475474, 00802526475771, 00802526480652, 00802526486944, 00802526497643, 00802526517174, 00802526526817, 00802526527623, 00802526557019, 00802526565304, 00802526579929, 00802526583803, 00802526602931, 00802526605468, 00802526606304, 00802526612411, 00802526615627, 00802526442827, 00802526444807, 00802526469374, 00802526477683, 00802526482267, 00802526497650, 00802526517198, 00802526540400, 00802526555169, 00802526564826, 00802526582325, 00802526585104, 00802526586316, 00802526586316, 00802526603013, 00802526606328, 00802526612435, 00802526615429, 00802526474057, 00802526477522, 00802526479649, 00802526482274, 00802526486470, 00802526497667, 00802526517211, 00802526540509, 00802526555206, 00802526564833, 00802526582318, 00802526585111, 00802526586408, 00802526603006, 00802526605420, 00802526606335, 00802526612404, 00802526615504, 00802526480669, 00802526605451, 00802526496653, 00802526526824, 00802526527630, 00802526527630, 00802526564840, 00802526581038, 00802526582035, 00802526584602, 00802526582035, 00802526605451, 00802526606403, and 00802526612459.
Where was it sold?
Distribution was nationwide, including Guam, Puerto Rico, and the U.S. Virgin Islands. There was also government/military distribution. Foreign distribution was made to Andorra, Argentina, Aruba, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, British Virgin Islands, Bulgaria, Cameroon, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Faroe Islands, Finland, France, French Polynesia, Germany, Great Britain, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Madagascar, Malaysia, Malta, Mauritius, Mexico, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Yemen, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan R.O.C., Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, Uruguay, Venezuela, Vietnam, and Zimbabwe.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Boston Scientific
All 626| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS I | Enroute Transcarotid Neuroprotection System recalled | Boston Scientific | Other | Nationwide |
| Device | CLASS I | Enroute Transcarotid Neuroprotection System Plus recalled | Boston Scientific | Other | Nationwide |
Other Pacemakers & Defibrillators recalls
All 494| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DermaV Laser System recalled by Lutronic | Lutronic | Other | Nationwide |
| Device | CLASS II | Convenience kits containing select SKUs of Polycarbonate recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits: Loop Pack DYNJ61460C Open Heart recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Medline convenience kits: Acl Pack DYNJ64161C Arthroscopy recalled | Medline Industries | Foreign Material | Nationwide |
| Device | CLASS II | Medline convenience kits: Bib AV Fistula Pack DYNJ58304B recalled | Medline Industries | Device Malfunction | Nationwide |