CLASS II Device recall · Reported 19 Dec 2012 · FDA Enforcement Report

Bronchoscope (flexible or rigid) and accessories recalled by ConMed

ConMed is recalling Bronchoscope (flexible or rigid) and accessories, distributed in 14 states. The recall was initiated on 20 Nov 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesBegining Lot 1207251, Ending Lot Code 1207251; Beginning Lot Code 1209171, Ending Lot Code 1210151. Lot codes on boxes and packaging contain a lot code in the following form 12 09 17 1 where 12 represents the year (2012), 09 represents the month (September), 17 is the day of the month, and 1 is the Manufacturing Shift Code (1, 2, 3, 4, 5, 6 or 9).
Quantity recalledTotal 452

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0502-2013
FDA event ID
63728
Recalling firm
ConMed, Utica, NY
Classification
Class II
Status
Terminated
Recall initiated
20 Nov 2012
FDA classified
10 Dec 2012
Reported
19 Dec 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

20 Nov 2012Recall initiated by company
10 Dec 2012Classified by FDA+20 days
19 Dec 2012Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles would not insert properly through the instrument channel of a 2.0 mm bronchoscope. If the device will not fit through the instrument channel of a 2.0 mm bronchoscope, it must be replaced with an alternate device and could result in minor inconvenience to the user.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

WANG Transbronchial Aspiration Needles Cat. No. MW-222, WANG Cytology Needles, Central or Peripheral Regions, 22 g x 13 mm long needle. --- Common Name: Bronchoscope (flexible or rigid) and accessories. The ConMed WANG Transbronchial Aspiration Needles are designed for use with a bronchoscope. They are used to puncture the tracheobronchial wall and to aspirate sufficient tissue and/or cell specimens to stage bronchogenic carcinoma. These are disposable sterile, single-use devices.

Where it was distributed

Worldwide Distribution - USA including CA , FL, IA, IN, MA, MD, ME, MO, MS, NC, NM, OH, PA, TX. Foreign distribution to Spain, Portugal, Japan and Hungary.

CaliforniaFloridaIowaIndianaMassachusettsMarylandMaineMissouriMississippiNorth CarolinaNew MexicoOhioPennsylvaniaTexas

Questions about this recall

Is Bronchoscope (flexible or rigid) and accessories recalled?

Yes. ConMed recalled Bronchoscope (flexible or rigid) and accessories on 20 Nov 2012. The recall is Class II and its status is Terminated. Recall number Z-0502-2013.

Why was it recalled?

ConMed received complaints from a user facility reporting that the WANG Transbronchial Aspiration Needles would not insert properly through the instrument channel of a 2.0 mm bronchoscope. If the device will not fit through the instrument channel of a 2.0 mm bronchoscope, it must be replaced with an alternate device and could result in minor inconvenience to the user.

Which lots are affected?

Begining Lot 1207251, Ending Lot Code 1207251; Beginning Lot Code 1209171, Ending Lot Code 1210151. Lot codes on boxes and packaging contain a lot code in the following form 12 09 17 1 where 12 represents the year (2012), 09 represents the month (September), 17 is the day of the month, and 1 is the Manufacturing Shift Code (1, 2, 3, 4, 5, 6 or 9).

Where was it sold?

Worldwide Distribution - USA including CA , FL, IA, IN, MA, MD, ME, MO, MS, NC, NM, OH, PA, TX. Foreign distribution to Spain, Portugal, Japan and Hungary.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from ConMed

All 92

Other Syringes & Needles recalls

All 1,060