CLASS II ONGOING Device recall · Reported 12 Aug 2020 · FDA Enforcement Report
Captivator Polypectomy Snare recalled by Boston Scientific
Boston Scientific is recalling Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire. The recall was initiated on 6 Jul 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2761-2020
- FDA event ID
- 86004
- Recalling firm
- Boston Scientific, Marlborough, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 6 Jul 2020
- FDA classified
- 5 Aug 2020
- Reported
- 12 Aug 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Surgical Instruments
Timeline
| 6 Jul 2020 | Recall initiated by company | |
| 5 Aug 2020 | Classified by FDA | +30 days |
| 12 Aug 2020 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.
Where it was distributed
World wide distribution.
Questions about this recall
Is Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire recalled?
Yes. Boston Scientific recalled Captivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire on 6 Jul 2020. The recall is Class II and its status is Ongoing. Recall number Z-2761-2020.
Why was it recalled?
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
Which lots are affected?
Outer Package UPN #: M00562321 Inner Package UPN #: M00562320 GTIN: 08714729019336 Lot #s: 25158564, 25245733, 25245735, 25330549, 25403001, 25439073, 25513671 Expiration Dates: 4-Feb-23 through 14-May-23
Where was it sold?
World wide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Boston Scientific
All 626| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS I | Enroute Transcarotid Neuroprotection System recalled | Boston Scientific | Other | Nationwide |
| Device | CLASS I | Enroute Transcarotid Neuroprotection System Plus recalled | Boston Scientific | Other | Nationwide |
Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Sterility | Nationwide |