CLASS I ONGOING Device recall · Reported 15 Mar 2023 · FDA Enforcement Report

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP) recalled

Datascope is recalling Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (, distributed nationwide in the US. The recall was initiated on 7 Feb 2023 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesAll serial numbers. Model Number 0998-00-0800-75, UDI-DI 10607567112312; Model Number 0998-00-0800-83, UDI-DI 10607567108407; Model Number 0998-00-0800-85, UDI-DI 10607567113449
Quantity recalled4062 Total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1146-2023
FDA event ID
91602
Recalling firm
Datascope, Mahwah, NJ
Classification
Class I
Status
Ongoing
Recall initiated
7 Feb 2023
FDA classified
9 Mar 2023
Reported
15 Mar 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Feb 2023Recall initiated by company
9 Mar 2023Classified by FDA+30 days
15 Mar 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Where it was distributed

Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.

Nationwide

Questions about this recall

Why was it recalled?

An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.

Which lots are affected?

All serial numbers. Model Number 0998-00-0800-75, UDI-DI 10607567112312; Model Number 0998-00-0800-83, UDI-DI 10607567108407; Model Number 0998-00-0800-85, UDI-DI 10607567113449

Where was it sold?

Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 91602
ReportedClassRecallCompanyReasonWhere
CLASS ICardiosave Hybrid Intra-Aortic Balloon Pumps (IABP) recalledDatascopeDevice MalfunctionNationwide

More recalls from Datascope

All 134
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIIBattery Charging Station recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IICS100 IABP recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IICS300 IABP recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IIICardiosave Hybrid recalled by DatascopeDatascopeOtherNationwide
DeviceCLASS IIICardiosave Rescue recalled by DatascopeDatascopeOtherNationwide

Other Heart & Vascular Devices recalls