CLASS I ONGOING Device recall · Reported 2 Oct 2024 · FDA Enforcement Report

Cardiosave Rescue Intra-aortic Balloon Pump (iabp) recalled by Datascope

Datascope is recalling Cardiosave Rescue Intra-aortic Balloon Pump (iabp), distributed nationwide in the US. The recall was initiated on 8 Aug 2024 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesProduct No. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85; UDI: 10607567112312, 10607567108407, 10607567113449; Lot No. ALL LOTS.
Quantity recalled5,475 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3135-2024
FDA event ID
95150
Recalling firm
Datascope, Mahwah, NJ
Classification
Class I
Status
Ongoing
Recall initiated
8 Aug 2024
FDA classified
20 Sep 2024
Reported
2 Oct 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

8 Aug 2024Recall initiated by company
20 Sep 2024Classified by FDA+43 days
2 Oct 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.

Where it was distributed

International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, IRAQ, IRAN, ITALY, JORDAN, JAPAN, SOUTH KOREA, KUWAIT, LEBANON, LITHUANIA, LIBYA, MOROCCO, MYANMAR, MEXICO, MALAYSIA, NETHERLANDS, NORWAY, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PALESTINE, PORTUGAL, SERBIA, RUSSIAN FEDERATION, SAUDI ARABIA, SWEDEN, SINGAPORE, SLOVAKIA, THAILAND, TURKEY, TAIWAN, UNITED KINGDOM, VIETNAM, YEMEN.

Nationwide

Questions about this recall

Is Cardiosave Rescue Intra-aortic Balloon Pump (iabp) recalled?

Yes. Datascope recalled Cardiosave Rescue Intra-aortic Balloon Pump (iabp) on 8 Aug 2024. The recall is Class I and its status is Ongoing. Recall number Z-3135-2024.

Why was it recalled?

Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.

Which lots are affected?

Product No. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85; UDI: 10607567112312, 10607567108407, 10607567113449; Lot No. ALL LOTS.

Where was it sold?

International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, IRAQ, IRAN, ITALY, JORDAN, JAPAN, SOUTH KOREA, KUWAIT, LEBANON, LITHUANIA, LIBYA, MOROCCO, MYANMAR, MEXICO, MALAYSIA, NETHERLANDS, NORWAY, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PALESTINE, PORTUGAL, SERBIA, RUSSIAN FEDERATION, SAUDI ARABIA, SWEDEN, SINGAPORE, SLOVAKIA, THAILAND, TURKEY, TAIWAN, UNITED KINGDOM, VIETNAM, YEMEN.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Datascope

All 134
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIIBattery Charging Station recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IICS100 IABP recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IICS300 IABP recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IIICardiosave Hybrid recalled by DatascopeDatascopeOtherNationwide
DeviceCLASS IIICardiosave Rescue recalled by DatascopeDatascopeOtherNationwide

Other Heart & Vascular Devices recalls

All 6,199