CLASS I ONGOING Device recall · Reported 28 Aug 2024 · FDA Enforcement Report

ClotTriever XL Catheter recalled by Inari Medical - Oak Canyon

Inari Medical - Oak Canyon is recalling ClotTriever XL Catheter, distributed nationwide in the US. The recall was initiated on 19 Jul 2024 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesUDI-DI:00850291007277. All lots. IU-01011 REV. C.
Quantity recalled2488

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2587-2024
FDA event ID
95069
Recalling firm
Inari Medical - Oak Canyon, Irvine, CA
Classification
Class I
Status
Ongoing
Recall initiated
19 Jul 2024
FDA classified
21 Aug 2024
Reported
28 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

19 Jul 2024Recall initiated by company
21 Aug 2024Classified by FDA+33 days
28 Aug 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.

Product as listed by the FDA

ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Thrombectomy System, Reference: IU-01011

Where it was distributed

Worldwide - US Nationwide distribution including in the states of PA, FL, GA, CA, IL, SC, NY, MN, UT, TN, WI, MI, KS, SD, AZ, AR, IN, KY, TX, AL, MS, DE, MA, OH, MO, WV, VA, NC, NV, CO, NE, OR, NM, NH, NJ, CT, ID, IA, WA, MD, OK, RI, DC, PR, LA, ND, VT and the countries of New Zealand, Chile, Australia.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichigan

Questions about this recall

Why was it recalled?

Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.

Which lots are affected?

UDI-DI:00850291007277. All lots. IU-01011 REV. C.

Where was it sold?

Worldwide - US Nationwide distribution including in the states of PA, FL, GA, CA, IL, SC, NY, MN, UT, TN, WI, MI, KS, SD, AZ, AR, IN, KY, TX, AL, MS, DE, MA, OH, MO, WV, VA, NC, NV, CO, NE, OR, NM, NH, NJ, CT, ID, IA, WA, MD, OK, RI, DC, PR, LA, ND, VT and the countries of New Zealand, Chile, Australia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Inari Medical - Oak Canyon

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIArtix MT Thrombectomy Device recalled by Inari Medical - Oak CanyonInari Medical - Oak CanyonOtherNationwide

Other Heart & Vascular Devices recalls