CLASS II Device recall · Reported 11 Mar 2020 · FDA Enforcement Report

Connected OR Cart,120 V recalled by Stryker

Stryker is recalling Stryker Connected OR Cart,120 V, distributed nationwide in the US. The recall was initiated on 23 Apr 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 07613327462784 Lot Numbers: 2019AN00005, 2019AN 00007, 2019AN00017, 2019AN00039, 2019AN00041, 2019AN00046, 2019AN00048, 2019AN00054, 2019AN00061, 2019AN00067, 2019AN00085, 2019AN00087
Quantity recalled12 Devices

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1445-2020
FDA event ID
84715
Recalling firm
Stryker, San Jose, CA
Classification
Class II
Status
Terminated
Recall initiated
23 Apr 2019
FDA classified
5 Mar 2020
Reported
11 Mar 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals

Timeline

23 Apr 2019Recall initiated by company
5 Mar 2020Classified by FDA+317 days
11 Mar 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential problem with firm s Connected OR Cart, 120V resulting in cart monitor not attaching flush with their monitor stow handle. The product is not manufactured to specifications and is positioned on the incorrect side that may lead to monitor falling from the monitor arm and a risk of user injury.

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

Stryker Connected OR Cart,120 V Ref Cat. #240-099-155, Qty 1, Rx Only, Manufactured for: Stryker, 5900 Optical Court, San Jose, CA 95138, UDI: 07613327462784 - Product Usage: The Connect OR Cart is a mobile, medical-grade cart intended to store and transport components of minimally invasive and open-field imaging systems such as the Stryker Advanced Imaging Modality (AIM) System, PINPOINT Endoscopic Fluorescence Imaging System, or SPY Portable Handled Imaging (SPY-PHI) System) during intraoperative use. The Cart also provides power, video, and data connections to the components.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of GA, TX and countries of Canada, Singapore, Thailand, Malaysia. No US Govt.

Nationwide GeorgiaTexas

Questions about this recall

Is Connected OR Cart,120 V recalled?

Yes. Stryker recalled Connected OR Cart,120 V on 23 Apr 2019. The recall is Class II and its status is Terminated. Recall number Z-1445-2020.

Why was it recalled?

There is a potential problem with firm s Connected OR Cart, 120V resulting in cart monitor not attaching flush with their monitor stow handle. The product is not manufactured to specifications and is positioned on the incorrect side that may lead to monitor falling from the monitor arm and a risk of user injury.

Which lots are affected?

UDI: 07613327462784 Lot Numbers: 2019AN00005, 2019AN 00007, 2019AN00017, 2019AN00039, 2019AN00041, 2019AN00046, 2019AN00048, 2019AN00054, 2019AN00061, 2019AN00067, 2019AN00085, 2019AN00087

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of GA, TX and countries of Canada, Singapore, Thailand, Malaysia. No US Govt.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.