CLASS II ONGOING Device recall · Reported 13 Dec 2023 · FDA Enforcement Report
Cooper Surgical H/S Elliptosphere Cath Set recalled
CooperSurgical is recalling Cooper Surgical H/S Elliptosphere Cath Set-For administering contrast media during, distributed nationwide in the US. The recall was initiated on 31 Mar 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0503-2024
- FDA event ID
- 93239
- Recalling firm
- CooperSurgical, Trumbull, CT
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 31 Mar 2022
- FDA classified
- 6 Dec 2023
- Reported
- 13 Dec 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Surgical Instruments
Timeline
| 31 Mar 2022 | Recall initiated by company | |
| 6 Dec 2023 | Classified by FDA | +615 days |
| 13 Dec 2023 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
Cooper Surgical H/S ELLIPTOSPHERE CATH 5 FR Set-For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes Model: 61-4005
Where it was distributed
Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLAND
Questions about this recall
Why was it recalled?
Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
Which lots are affected?
UDI-DI:60888937016446 S/N: 313121, 313568
Where was it sold?
Nationwide Foreign: BELGIUM CANADA MALAYSIA NETHERLANDS SPAIN SWEDEN SWITZERLAND
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 93239This product is part of a recall event; the main page for the event is Cooper Surgical H/S Elliptosphere Cath -For administering recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Cooper Surgical H/S Elliptosphere Cath -For administering recalled | CooperSurgical | Packaging Defects | Nationwide | |
| CLASS II | H/S Catheter SET FR 5 with Integated Stylet recalled by CooperSurgical | CooperSurgical | Packaging Defects | Nationwide |
More recalls from CooperSurgical
All 86| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS II | HSG Procedure Tray recalled by CooperSurgical | CooperSurgical | Sterility | Nationwide |
| Device | CLASS II | Endosee System Convenience Kit recalled by CooperSurgical | CooperSurgical | Sterility | Nationwide |
| Device | CLASS II | Endosee System Convenience Kit with IV Tube recalled by CooperSurgical | CooperSurgical | Sterility | Nationwide |
Other Surgical Instruments recalls
All 3,891| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Adaptable Beach Chair Beach Chair surgical patient positioner recalled | Quantum OPS | Device Malfunction | Nationwide |
| Device | CLASS II | Laparoscopic Prostatectomy kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Halyard Laparoscopy Comprehensive Pack II recalled | GRI Medical & Electronic Technology | Other | Nationwide |
| Device | CLASS II | Gyn Laparoscopy/Laparotomy kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Gyn Laparoscopy Pack kits recalled by Avid Medical | Avid Medical | Other | Nationwide |