CLASS I ONGOING Device recall · Reported 13 Sep 2017 · FDA Enforcement Report

CS 100i Intra-Aortic Balloon Pump recalled by Maquet Datascope Corp

Maquet Datascope Corp - Cardiac Assist Division is recalling CS 100i Intra-Aortic Balloon Pump, distributed nationwide in the US. The recall was initiated on 17 Jul 2017 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codes0998-UC-0446HXX; 0998-UC-0479HXX
Quantity recalled14,547 units (5,049 domestically & 9,498 units internationally)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3085-2017
FDA event ID
77877
Classification
Class I
Status
Ongoing
Recall initiated
17 Jul 2017
FDA classified
5 Sep 2017
Reported
13 Sep 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

17 Jul 2017Recall initiated by company
5 Sep 2017Classified by FDA+50 days
13 Sep 2017Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

CS 100i Intra-Aortic Balloon Pump. It supports the heart's left ventricle by increasing coronary perfusion and reducing left ventricular work. Coronary perfusion is increased by augmenting blood pressure during the diastolic phase of the cardiac cycle. This increase in aortic pressure promotes more blood flow through the coronary arteries. Left ventricular work is reduced by decreasing aortic end diastolic pressure and reducing resistance to ventricular ejection, resulting in a decrease in blood pressure during the systolic phase of the cardiac cycle.

Where it was distributed

Worldwide Distribution - US Nationwide International: AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MALDIVES, MALTA, MAURITIUS, MEXICO, MOROCCO, MYANMAR, NEPAL, NETHERLANDS, NEW ZEALAND, NORTH KOREA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD, TUNISIA, TURKEY, TURKMENISTAN, UGANDA, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM & YEMEN.

Nationwide Georgia

Questions about this recall

Why was it recalled?

False blood back detection alarm and the ingress of fluids into the IABP affecting various electronic circuit boards could potentially prevent initiation or continuation of therapy.

Which lots are affected?

0998-UC-0446HXX; 0998-UC-0479HXX

Where was it sold?

Worldwide Distribution - US Nationwide International: AFGHANISTAN, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MALDIVES, MALTA, MAURITIUS, MEXICO, MOROCCO, MYANMAR, NEPAL, NETHERLANDS, NEW ZEALAND, NORTH KOREA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD, TUNISIA, TURKEY, TURKMENISTAN, UGANDA, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, VENEZUELA, VIETNAM & YEMEN.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 77877

This product is part of a recall event; the main page for the event is CS 100 Intra-Aortic Balloon Pump recalled by Maquet Datascope Corp.

More recalls from Maquet Datascope Corp - Cardiac Assist Division

All 14