CLASS II ONGOING Device recall · Reported 1 Apr 2026 · FDA Enforcement Report

Custom Abutment AS Ti, for Straumann RC, incl recalled

Straumann USA is recalling Custom Abutment AS Ti, for Straumann RC, incl, distributed nationwide in the US. The recall was initiated on 17 Mar 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00;
Quantity recalled7 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1633-2026
FDA event ID
98565
Recalling firm
Straumann USA, Andover, MA
Classification
Class II
Status
Ongoing
Recall initiated
17 Mar 2026
FDA classified
24 Mar 2026
Reported
1 Apr 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

17 Mar 2026Recall initiated by company
24 Mar 2026Classified by FDA+7 days
1 Apr 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;

Where it was distributed

Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.

Nationwide CaliforniaColoradoFloridaIowaIllinoisMinnesotaMissouriMontanaOregonWashington

Questions about this recall

Why was it recalled?

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

Which lots are affected?

Catalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00;

Where was it sold?

Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 98565

This product is part of a recall event; the main page for the event is Custom Abutment Ti, for Straumann RC, incl recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IICustom Abutment Ti, for Straumann RC, incl recalledStraumann USAPackaging DefectsNationwide

More recalls from Straumann USA

All 22

Other Other Medical Devices recalls

All 13,884