CLASS II Device recall · Reported 19 Mar 2014 · FDA Enforcement Report

CWP Reverse Osmosis System recalled by Mar Cor Purification

Mar Cor Purification is recalling Mar Cor Purification CWP Reverse Osmosis System, distributed nationwide in the US. The recall was initiated on 27 Jun 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesserial numbers: 1289064, 1289069, 1289159, 1289259, 1289387, 1289650, 1289747, 1289826, 1289949, 1290088, 1290151, 1290333, 1290367, 1290480, 1290489, 1290845, 1290988, 1291042, 1291120, 1291233, 1291373, 1291378, 1291634, 1291777, 1291828, 1291933, 1292031, 1292114, 1292185, 1292185, 1292192, 1292391, 1292546, 1292597, 1292670, 1292674, 1292899, 1292960, 1293010, 1293039, 1293185, 1293190, 1293311, 1293379, 1293414, 1293687, 1293724, 1293771, 1293777, 1293953, 1293957, 1294093, 1294103, 1294171, 1294307, 1294362, 1294433, 1294516, 1294589, 1294727, 1294864, 1294999, 1295113, 1295137, 1295160, 1295186, 1295274, 1295391, 1295514, 1295622, 1295813, 1295871, 1296237, 1296241, 1296246, 1296484, 1296491, 1296582, 1296621, 1296650, 1296695, 1296806, 1297041, 1297046, 1297118, 1297231, 1297234, 1297372, 1297375, 1297394, 1297453, 1297484, 1297610, 1297655, 1297939, 1297943, 1298037, 1298210, 1298266, 1298270, 1298377, 1298407, 1298450, 1298457, 1298575, 1298619, 1298624, 1298704, 1298734, 1298797, 1298820, 1298879, 1298913, 1298952, 1298953, 1299037, 1299120, 1299144, 1299170, 1299241, 1299262, 1299285, 1299331, 1299393, 1299473, 1299512, 1299599, 1299655, 1299656, 1299732.
Quantity recalled130

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1201-2014
FDA event ID
67516
Recalling firm
Mar Cor Purification, Minneapolis, MN
Classification
Class II
Status
Terminated
Recall initiated
27 Jun 2012
FDA classified
13 Mar 2014
Reported
19 Mar 2014
Type
Voluntary: Firm initiated
Customers notified by
Telephone
Reason category
Foreign Material

Timeline

27 Jun 2012Recall initiated by company
13 Mar 2014Classified by FDA+624 days
19 Mar 2014Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.

Where it was distributed

nationwide. No distribution outside the US.

Nationwide

Questions about this recall

Is CWP Reverse Osmosis System recalled?

Yes. Mar Cor Purification recalled CWP Reverse Osmosis System on 27 Jun 2012. The recall is Class II and its status is Terminated. Recall number Z-1201-2014.

Why was it recalled?

There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.

Which lots are affected?

serial numbers: 1289064, 1289069, 1289159, 1289259, 1289387, 1289650, 1289747, 1289826, 1289949, 1290088, 1290151, 1290333, 1290367, 1290480, 1290489, 1290845, 1290988, 1291042, 1291120, 1291233, 1291373, 1291378, 1291634, 1291777, 1291828, 1291933, 1292031, 1292114, 1292185, 1292185, 1292192, 1292391, 1292546, 1292597, 1292670, 1292674, 1292899, 1292960, 1293010, 1293039, 1293185, 1293190, 1293311, 1293379, 1293414, 1293687, 1293724, 1293771, 1293777, 1293953, 1293957, 1294093, 1294103, 1294171, 1294307, 1294362, 1294433, 1294516, 1294589, 1294727, 1294864, 1294999, 1295113, 1295137, 1295160, 1295186, 1295274, 1295391, 1295514, 1295622, 1295813, 1295871, 1296237, 1296241, 1296246, 1296484, 1296491, 1296582, 1296621, 1296650, 1296695, 1296806, 1297041, 1297046, 1297118, 1297231, 1297234, 1297372, 1297375, 1297394, 1297453, 1297484, 1297610, 1297655, 1297939, 1297943, 1298037, 1298210, 1298266, 1298270, 1298377, 1298407, 1298450, 1298457, 1298575, 1298619, 1298624, 1298704, 1298734, 1298797, 1298820, 1298879, 1298913, 1298952, 1298953, 1299037, 1299120, 1299144, 1299170, 1299241, 1299262, 1299285, 1299331, 1299393, 1299473, 1299512, 1299599, 1299655, 1299656, 1299732.

Where was it sold?

nationwide. No distribution outside the US.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Mar Cor Purification

All 3

Other Other Medical Devices recalls

All 13,778