CLASS II ONGOING Device recall · Reported 8 Jul 2026 · FDA Enforcement Report

Cyberknife Treatment Deliver System equipped recalled

Accuray is recalling Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer, distributed nationwide in the US. The recall was initiated on 16 Jun 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll systems prior with software version 11X that utilize the Exchange Table
Quantity recalled48 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2640-2026
FDA event ID
99075
Recalling firm
Accuray, Madison, WI
Classification
Class II
Status
Ongoing
Recall initiated
16 Jun 2026
FDA classified
1 Jul 2026
Reported
8 Jul 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

16 Jun 2026Recall initiated by company
1 Jul 2026Classified by FDA+15 days
8 Jul 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

Where it was distributed

Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.

Nationwide

Questions about this recall

Why was it recalled?

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

Which lots are affected?

All systems prior with software version 11X that utilize the Exchange Table

Where was it sold?

Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Accuray

All 22

Other Masks, Gloves & PPE recalls

All 450