CLASS I ONGOING Device recall · Reported 24 Jun 2026 · FDA Enforcement Report
Description/REF: PI Agba Picc KIT recalled by Arrow International
Arrow International is recalling Description/REF: PI Agba Picc KIT: 3L 6 FR X 40 CM W/VPS/Ask-44063-NS, distributed nationwide in the US. The recall was initiated on 21 Apr 2026 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2370-2026
- FDA event ID
- 98793
- Recalling firm
- Arrow International, Morrisville, NC
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 21 Apr 2026
- FDA classified
- 12 Jun 2026
- Reported
- 24 Jun 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Other Medical Devices
Timeline
| 21 Apr 2026 | Recall initiated by company | |
| 12 Jun 2026 | Classified by FDA | +52 days |
| 24 Jun 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT: 2L 5.5 FR X 55 CM CG+ VPS/ASK-45552-RH AGBA PICC NaviCurve: 2L 5.5FR x 55CM BP/DLX-45552-CURVB AGBA PICC NaviCurve: 2L 5.5FR x 55CM TCG/DLX-45552-CURVC AGBA PICC: 2L 5.5FR x 55CM w Biopatch/DLX-45552-HPKB AGBA PICC NaviCurve: 3L 6FR x 55CM BP/DLX-45563-CURVB PI PICC KIT: 4 FR X 50 CM/ASK-35041-NH PI PICC KIT: 2-L 5 FR X 50 CM/ASK-35052-NH PI PICC G4 Stylet: 1L 4FR x 55CM w CHG/DLX-35541-VPSC PI PICC NaviCurve: 2L 5FR x 55CM w BP/DLX-35552-CURVB PI PICC NaviCurve: 3L 6FR x 55CM w BP/DLX-35563-CURVB PI PICC G4 Stylet: 3L 6FR x 55CM w BP/DLX-35563-VPSB AGBA PI PICC KIT: 2L 5.5 FR X 55CM/CDC-45552-HPK1A PI AGBA PICC KIT: 3-L 6 FR X 55 CM/CDC-45563-HPK1A PI AGBA PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UOI1 PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-VFH1 PI AGBA PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-UOI1 PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-VFH1 PI AGBA PICC KIT: 3L 6 FR X 55 CM W/VPS/ASK-45563-UWH1 AGBA PICC NaviCurve: 1L 4.5FR x 55CM BP/DLX-45541-CURVB AGBA PICC NaviCurve: 1L 4.5FR x 55CM TCG/DLX-45541-CURVC AGBA PICC G4 Stylet: 1L 4.5FR x 55CM TCG/DLX-45541-VPSC AGBA PICC G4 Stylet: 2L 5.5FR x 55CM BP/DLX-45552-VPSB AGBA PICC G4 Stylet: 2L 5.5FR x 55CM TCG/DLX-45552-VPSC PI PICC:1L 4FR X 50CM PRELOAD VPS STYLET/CDC-35041-VPS PICC KIT: 3 FR X 55 CM/CDC-05531-PK1A
Where it was distributed
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Which lots are affected?
REF(Material)/UDI-DI/Lot(Batch): ASK-44063-NS/10801902193275/33F25F0100; ASK-45052-NS/10801902193275/33F24L0864 ASK-45541-MHBV/10801902193275, 10801902167696/33F25D0410, 33F26C0012, 33S24A0690, 33F25B0574; ASK-45552-RH/10801902193275/33F25B0575, 33F25D0361, 33F25F0197, 33F25K0373, 33F25A0082; DLX-45552-CURVB/10801902193275, 10801902208702/33F25E0506, 33F25E0833, 33F26C0038, 33S24B0518, 33F24L0982; DLX-45552-CURVC/10801902193275, 10801902208733/33F25B0386, 33F25D0381, 33F25D0624, 33F25E0765, 33F25F0225, 33F25H0602, 33F26C0182, 33F25B0390; DLX-45552-HPKB/10801902193275, 10801902208825, 10801902208856/33F25D0454, 33F25D0677, 33F25E0834, 33F25J0446, 33F25M0172, 33F26B0867, 33S24A0586, 33S24A1370, 33F24F0097; DLX-45563-CURVB/10801902193275/33F25C0339, 33F25E0836, 33F25A0321; ASK-35041-NH/10801902221077, 10801902193275/33F26B0140 33F25G0460; ASK-35052-NH/10801902193275/33F25H0412, 33F24L1080; DLX-35541-VPSC/10801902193275/33F25C0449, 33F25D0245, 33F25H0501, 33F24L1148; DLX-35552-CURVB/10801902193275, 10801902208764/33F25C0214, 33F25E0511 ,33F25E0832 ,33F26C0120 33F26C0732, 33F25C0373; DLX-35563-CURVB/10801902193275, 10801902208771/33F25C0337, 33F26B0155, 33F25D0453; DLX-35563-VPSB/10801902193275/33F25C0395; CDC-45552-HPK1A/10801902193275, 10801902188462/33F25D0067, 33F25D0243, 33F25E0452, 33F25E0658, 33F25E0829, 33F26B0891, 33F26C0202, 33S23K0168, 33S24A0419, 33S24A0523, 33S24A1387, 33F25K0317; CDC-45563-HPK1A/10801902193275, 10801902188479/33F25G0283, 33F25H0626, 33S24A0543, 33F24D0497; ASK-45541-UOI1/10801902193275, 10801902221664/33F25H0411, 33F25J0399, 33S24A1368, 33F26A0226; ASK-45541-VFH1/10801902193275, 10801902218886/33F25B0655, 33F25E0573, 33F25G0210, 33F26A1533, 33F26B0841, 33F25G0584; ASK-45552-UOI1/10801902193275/33F26A0227; ASK-45552-VFH1/10801902193275/33F25D0362, 33F25E0866, 33F25J0226, 33F25M0180, 33F26A1455, 33F25H0408; ASK-45563-UWH1/10801902193275/33F25D0660, 33F25J0227, 33F24K0066; DLX-45541-CURVB/10801902193275/33F25B0385, 33F25D0250, 33F25E0760, 33F25J0336, 33F25K0194, 33F24L1115; DLX-45541-CURVC/10801902193275/33F25B0507, 33F25C0216, 33F25E0761, 33F25F0223, 33F25G0362, 33F25J0337, 33F25E0762; DLX-45541-VPSC/10801902193275, 10801902208610/33F24L0390, 33F24L1130, 33F25C0247, 33F25D0251, 33F25D0380, 33F25D0623, 33F25E0764, 33F25E0900, 33F25F0224, 33F25H0604, 33F25J0445, 33F26A1541, 33F26B0460, 33F26C0220, 33F26C0717, 33F25B0387; DLX-45552-VPSB/10801902193275/33F25C0249, 33F25D0253, 33F25D0314, 33F25D0382, 33F24L0392; DLX-45552-VPSC/10801902193275, 10801902208634/33F24L0939, 33F25B0508, 33F25D0254, 33F25D0315, 33F25D0383,33F25D0626, 33F25E0581, 33F25E0709, 33F25E0835, 33F25J0129, 33F25J0257, 33F25M0167, 33F26B0712, 33F26B0937, 33F25B0388; CDC-35041-VPS/10801902186741, 10801902193275/33X25K0001, 33F25E0425; CDC-05531-PK1A/10801902193275, 60801902123420/33F25C0641, 33F25D0435, 33F25E0646, 33F25J0416, 33F26B0711, 33F25D0765
Where was it sold?
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
34 other products · FDA event 98793This product is part of a recall event; the main page for the event is Description/REF: PI Jacc KIT: 2L 5.5 FR X 15 CM CG+/Ask-41552-Ltac recalled.
More recalls from Arrow International
All 525| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Description/REF: Agba Picc G4 Stylet: 3L 6FR x 55CM recalled | Arrow International | Sterility | Nationwide |
| Device | CLASS I | Description/REF: Hemodialysis KIT: 2-Lumen 8 FR X 11 CM/AK recalled | Arrow International | Sterility | Nationwide |
| Device | CLASS I | Description/REF: PI CVC KIT recalled by Arrow International | Arrow International | Sterility | Nationwide |
| Device | CLASS I | Description/REF: PI Picc: 3: 6FR x 55CM w CHG/ Dlx-35563-Hpkc; Agba recalled | Arrow International | Sterility | Nationwide |
| Device | CLASS I | Description/REF: Pacing PSI KIT: 6 FR/AI-06210-KS, Pacing PSI KIT recalled | Arrow International | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |