CLASS II ONGOING Device recall · Reported 7 Apr 2021 · FDA Enforcement Report

DiamondTemp Unidirectional Small Curve Ablation Catheter recalled

Epix Therapeutics is recalling DiamondTemp Unidirectional Small Curve Ablation Catheter, 7. The recall was initiated on 22 Feb 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll Lots
Quantity recalled100 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1317-2021
FDA event ID
87412
Recalling firm
Epix Therapeutics, Sunnyvale, CA
Classification
Class II
Status
Ongoing
Recall initiated
22 Feb 2021
FDA classified
29 Mar 2021
Reported
7 Apr 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

22 Feb 2021Recall initiated by company
29 Mar 2021Classified by FDA+35 days
7 Apr 2021Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

EPIX Therapeutics DiamondTemp Unidirectional Small Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF CEDT100S, SterileEO, UDI 00812499030198

Where it was distributed

Product was not distributed in the US. OUS: Germany, Spain, France, Belgium, United Kingdom, Slovenia, Italy, Canada

Questions about this recall

Why was it recalled?

It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli.

Which lots are affected?

All Lots

Where was it sold?

Product was not distributed in the US. OUS: Germany, Spain, France, Belgium, United Kingdom, Slovenia, Italy, Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

5 other products · FDA event 87412

This product is part of a recall event; the main page for the event is DiamondTemp Bidrectional Large Curve Ablation Catheter recalled.

More recalls from Epix Therapeutics

All 6

Other Heart & Vascular Devices recalls

All 6,199