CLASS II ONGOING Device recall · Reported 12 Aug 2026 · FDA Enforcement Report

Drive and munevo Drive r-net Powered Wheelchair Hands-Free recalled

MUNEVO is recalling munevo Drive and munevo Drive r-net Powered Wheelchair Hands-Free Control, distributed nationwide in the US. The recall was initiated on 19 Jun 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAffected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699
Quantity recalled19

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2856-2026
FDA event ID
99328
Recalling firm
MUNEVO, Munich
Classification
Class II
Status
Ongoing
Recall initiated
19 Jun 2026
FDA classified
31 Jul 2026
Reported
12 Aug 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

19 Jun 2026Recall initiated by company
31 Jul 2026Classified by FDA+42 days
12 Aug 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

Where it was distributed

Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.

Nationwide CaliforniaFloridaNorth CarolinaTexasUtahWashington

Questions about this recall

Is Drive and munevo Drive r-net Powered Wheelchair Hands-Free Control Application/Smart recalled?

Yes. MUNEVO recalled Drive and munevo Drive r-net Powered Wheelchair Hands-Free Control Application/Smart on 19 Jun 2026. The recall is Class II and its status is Ongoing. Recall number Z-2856-2026.

Why was it recalled?

A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.

Which lots are affected?

Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699

Where was it sold?

Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other Wheelchairs, Beds & Lifts recalls

All 172