CLASS II Device recall · Reported 2 May 2018 · FDA Enforcement Report
EEA" Hemorrhoid and Prolapse Stapler Set recalled by Covidien
Covidien is recalling EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology, Product Number, distributed nationwide in the US. The recall was initiated on 9 Apr 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1527-2018
- FDA event ID
- 79905
- Recalling firm
- Covidien, North Haven, CT
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 9 Apr 2018
- FDA classified
- 26 Apr 2018
- Reported
- 2 May 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Heart & Vascular Devices
Timeline
| 9 Apr 2018 | Recall initiated by company | |
| 26 Apr 2018 | Classified by FDA | +17 days |
| 2 May 2018 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There is a potential for improper welding of the yellow staple guide to the instrument. Use of a device with an improperly welded staple guide may result in improper staple formation potentially leading to bleeding or anastomotic leak.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology 3.5MM, Product Number HEM3335, GTIN Number 1 - 10884521047518, GTIN Number 2 - 20884521047515 Product Usage: The EEA Hemorrhoid and prolapse stapler set with DST Series Technology (instrument) is designed for use as a stapler instrument for control of rectal prolapse and hemorrhoid disease. The instrument places a circular, double-staggered row of titanium DST staples and removes a circular tissue specimen. The instrument is offered in a 33mm diameter size with either 3.5mm or 4.8mm staples. The accompanying accessories comprised of an access port, anoscope, and dilator. The access port aids the insertion and removal of the anoscope and the stapler device by dilating the anal sphincter throughout the procedure. The anoscope is designed to facilitate pursestring suture application above the dentate line. It is designed with a guide and markings to assist in creating a circumferential pursestring suture at a consistent placement within the rectum. The dilator is designed to aid the initial insertion of the anoscope and port by retracting the rectal tissue. Both the anoscope and port are transparent to facilitate visualization of the anal canal. The EEA Hemorrhoid and prolapse stapler set with DST Series Technology has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease. The EEA Hemorrhoid and prolapse stapler also has application in the distal alimentary tract for the creation of end-to-end and end-to-side anastomoses.
Where it was distributed
Worldwide Distribution - US Nationwide in the US states of : AL, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, SC, TN, TX, UT, and VA. The devices were distributed to the following foreign countries: UAE, Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Indonesia, Italy, Korea, Malaysia, Netherlands, Poland, Portugal, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Uganda, UK, and Vietnam.
Nationwide AlabamaCaliforniaFloridaGeorgiaIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiNorth CarolinaNebraskaNew JerseyNew York
Questions about this recall
Is EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology, Product Number recalled?
Yes. Covidien recalled EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology, Product Number on 9 Apr 2018. The recall is Class II and its status is Terminated. Recall number Z-1527-2018.
Why was it recalled?
There is a potential for improper welding of the yellow staple guide to the instrument. Use of a device with an improperly welded staple guide may result in improper staple formation potentially leading to bleeding or anastomotic leak.
Which lots are affected?
Lots: N7J1145MX, N7L0380MX, N7L0762MX, N7L1077MX, N7M0733MX,N7K0092MX, N7L0457MX, N7L0868MX, N7M0144MX, N7M0835MX, N7K0692MX, N7L0594MX, N7L0940MX, N7M0185MX N8A0144MX, N7K0693MX, N7L0676MX, N7L1076MX, N7M0732MX, N8A0166MX
Where was it sold?
Worldwide Distribution - US Nationwide in the US states of : AL, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, SC, TN, TX, UT, and VA. The devices were distributed to the following foreign countries: UAE, Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Indonesia, Italy, Korea, Malaysia, Netherlands, Poland, Portugal, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Uganda, UK, and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Covidien
All 493| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | EndoFLIP System that includes recalled by Covidien | Covidien | Other | Nationwide |
| Device | CLASS II | Dermalon 88861757-31 Dermalon* 4-0 Blu C13 recalled by Covidien | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Surgipro SP-623 Surgipro* 2-0 BLU SC2 X36 SP recalled by Covidien | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | V-Loc 90 Absorbable Wound Closure Device Vloc 90 0 VL recalled | Covidien | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Surgipro II S8698GX Surgipro II BL 5/0 18 P-13 S8698GX recalled | Covidien | Lead & Heavy Metals | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |