CLASS II ONGOING Device recall · Reported 5 Aug 2026 · FDA Enforcement Report

EOPA 3D Arterial Cannula, intended for use recalled

Medtronic Perfusion Systems is recalling EOPA 3D Arterial Cannula, intended for use with cardiopulmonary bypass as an arterial, distributed nationwide in the US. The recall was initiated on 24 Jun 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 20199150023251, Lot Numbers: 0232709184 0232745036 0232798416 0232862784 0232862787; UDI-DI: 00199150023257, Lot Numbers: 0232862787
Quantity recalled1,424 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2811-2026
FDA event ID
99320
Recalling firm
Medtronic Perfusion Systems, Brooklyn Park, MN
Classification
Class II
Status
Ongoing
Recall initiated
24 Jun 2026
FDA classified
27 Jul 2026
Reported
5 Aug 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

24 Jun 2026Recall initiated by company
27 Jul 2026Classified by FDA+33 days
5 Aug 2026Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

Which lots are affected?

UDI-DI: 20199150023251, Lot Numbers: 0232709184 0232745036 0232798416 0232862784 0232862787; UDI-DI: 00199150023257, Lot Numbers: 0232862787

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

5 other products · FDA event 99320

This product is part of a recall event; the main page for the event is EOPA Elongated One-Piece Arterial Cannula DIL TIP recalled.

More recalls from Medtronic Perfusion Systems

All 116

Other Heart & Vascular Devices recalls

All 6,199