CLASS II ONGOING Device recall · Reported 13 Nov 2024 · FDA Enforcement Report

epoc BGEM BUN Test Card for the epoc Blood Analysis System recalled

Siemens Healthcare Diagnostics is recalling epoc BGEM BUN Test Card for the epoc Blood Analysis System, SMN 10736515, distributed nationwide in the US. The recall was initiated on 24 Sep 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60
Quantity recalled1292 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0286-2025
FDA event ID
95504
Recalling firm
Siemens Healthcare Diagnostics, Norwood, MA
Classification
Class II
Status
Ongoing
Recall initiated
24 Sep 2024
FDA classified
1 Nov 2024
Reported
13 Nov 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

24 Sep 2024Recall initiated by company
1 Nov 2024Classified by FDA+38 days
13 Nov 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515

Where it was distributed

US Nationwide. Canada.

Nationwide

Questions about this recall

Why was it recalled?

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.

Which lots are affected?

UDI-DI: 00809708121860; Lot Numbers: 08-24092-60 08-24093-60 08-24094-60 08-24095-60 08-24096-60

Where was it sold?

US Nationwide. Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.