CLASS II ONGOING Device recall · Reported 6 May 2026 · FDA Enforcement Report

epoc BGEM BUN Test Card for the epoc Blood Analysis System recalled

Siemens Healthcare Diagnostics is recalling epoc BGEM BUN Test Card for the epoc Blood Analysis System, distributed nationwide in the US. The recall was initiated on 2 Apr 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSiemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);
Quantity recalled71725 (38025 US, 33700 OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2005-2026
FDA event ID
98680
Recalling firm
Siemens Healthcare Diagnostics, Norwood, MA
Classification
Class II
Status
Ongoing
Recall initiated
2 Apr 2026
FDA classified
29 Apr 2026
Reported
6 May 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

2 Apr 2026Recall initiated by company
29 Apr 2026Classified by FDA+27 days
6 May 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.

Nationwide

Questions about this recall

Why was it recalled?

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

Which lots are affected?

Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.