CLASS II ONGOING Device recall · Reported 3 Dec 2025 · FDA Enforcement Report

Evis Exera Iii Duodenovideoscope Olympus Tjf-Q190V recalled

Olympus Corporation of the Americas is recalling Evis Exera Iii Duodenovideoscope Olympus Tjf-Q190V, distributed nationwide in the US. The recall was initiated on 14 Oct 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel/Catalog Number: TJF-Q190V; UDI: 04953170405563 / 04953170452024; Serial Numbers: All;
Quantity recalled7023 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0579-2026
FDA event ID
97681
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA
Classification
Class II
Status
Ongoing
Recall initiated
14 Oct 2025
FDA classified
21 Nov 2025
Reported
3 Dec 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Processing Deviation

Timeline

14 Oct 2025Recall initiated by company
21 Nov 2025Classified by FDA+38 days
3 Dec 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.

Product as listed by the FDA

Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Continued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.

Which lots are affected?

Model/Catalog Number: TJF-Q190V; UDI: 04953170405563 / 04953170452024; Serial Numbers: All;

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Olympus Corporation of the Americas

All 423

Other Surgical Instruments recalls

All 3,891