CLASS II ONGOING Device recall · Reported 30 Sep 2026 · FDA Enforcement Report

Olympus BiCOAG Hemostasis Probe Coaxial Connector recalled

Olympus Corporation of the Americas is recalling Olympus BiCOAG Hemostasis Probe Coaxial Connector, distributed nationwide in the US. The recall was initiated on 19 Aug 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: Model: CD-B620LA. UDI-DI: 00821925039483. Lot Numbers: All unexpired lots.
Quantity recalled2,142 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3246-2026
FDA event ID
99609
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA
Classification
Class II
Status
Ongoing
Recall initiated
19 Aug 2026
FDA classified
18 Sep 2026
Reported
30 Sep 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

19 Aug 2026Recall initiated by company
18 Sep 2026Classified by FDA+30 days
30 Sep 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.

Product as listed by the FDA

Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA.

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, Latin America, Singapore, Hong Kong, and EU.

Nationwide

Questions about this recall

Why was it recalled?

Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.

Which lots are affected?

Lot Code: Model: CD-B620LA. UDI-DI: 00821925039483. Lot Numbers: All unexpired lots.

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Canada, Latin America, Singapore, Hong Kong, and EU.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 6 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 99609

This product is part of a recall event; the main page for the event is Olympus BiCOAG Hemostasis Probe Fixed Pin Connector recalled.

More recalls from Olympus Corporation of the Americas

All 420

Other Other Medical Devices recalls

All 13,857