CLASS II Device recall · Reported 13 May 2015 · FDA Enforcement Report

ExacTrac 6.x recalled by Brainlab AG

Brainlab is recalling ExacTrac 6.x. is, distributed nationwide in the US. The recall was initiated on 6 Mar 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware Versions: ExacTrac v. 6.x (v. 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0); Model/Catalog Numbers: 49971, CBCT IMPORT & ALIGNMENT SOFTWARE; 20836, EXACTRAC 6.0 VARIAN CBCT LICENSE; 48320, ET UPGRADE SOFTWARE 6.0 TO 6.1; 48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA; 49973B, ET DATA PREP / REVIEW SYSTEM; 49973A, ET DATA PREP / REVIEW SYSTEM; 49973, ET DATA PREP / REVIEW SYSTEM; 49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT; 48302, ET DATA PREP / REVIEW SYSTEM
Quantity recalled25 systems (US only); 53 systems (Foreign)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1582-2015
FDA event ID
70951
Recalling firm
Brainlab, Feldkirchen, N/A
Classification
Class II
Status
Terminated
Recall initiated
6 Mar 2015
FDA classified
6 May 2015
Reported
13 May 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

6 Mar 2015Recall initiated by company
6 May 2015Classified by FDA+61 days
13 May 2015Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

ExacTrac 6.x Patient Positioning System: Potentially incorrect patient positioning when using the ExacTrac Cone Beam CT (CBCT) with a CBCT acquired at a couch angle other than 0.0 degrees.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ExacTrac 6.x. is software used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures.

Where it was distributed

*** US: Nationwide; *** FOREIGN: Austria, China, Denmark, France, Germany, Hong Kong, Japan, Netherlands, Qatar, Russian Federation, South Korea, Switzerland

Nationwide

Questions about this recall

Why was it recalled?

ExacTrac 6.x Patient Positioning System: Potentially incorrect patient positioning when using the ExacTrac Cone Beam CT (CBCT) with a CBCT acquired at a couch angle other than 0.0 degrees.

Which lots are affected?

Software Versions: ExacTrac v. 6.x (v. 6.0.0, 6.0.1, 6.0.2, 6.0.3, 6.0.4, 6.0.5; and v. 6.1.0); Model/Catalog Numbers: 49971, CBCT IMPORT & ALIGNMENT SOFTWARE; 20836, EXACTRAC 6.0 VARIAN CBCT LICENSE; 48320, ET UPGRADE SOFTWARE 6.0 TO 6.1; 48320-99, ET UPGRADE SOFTWARE 6.0 TO 6.1 SLA; 49973B, ET DATA PREP / REVIEW SYSTEM; 49973A, ET DATA PREP / REVIEW SYSTEM; 49973, ET DATA PREP / REVIEW SYSTEM; 49998B, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 49998A, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 49998, ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT; 48304, ET 6.1 DATA PREP/REVIEW SYSTEM UPG. KIT; 48302, ET DATA PREP / REVIEW SYSTEM

Where was it sold?

*** US: Nationwide; *** FOREIGN: Austria, China, Denmark, France, Germany, Hong Kong, Japan, Netherlands, Qatar, Russian Federation, South Korea, Switzerland

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Brainlab

All 39

Other Medical Software recalls

All 932