CLASS II Device recall · Reported 21 Nov 2012 · FDA Enforcement Report

Fill Polymer Kit, Endovascular Graft recalled by Trivascular

Trivascular is recalling TriVascular Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European, distributed nationwide in the US. The recall was initiated on 19 Oct 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesThe following lot numbers are in distributor inventory in Europe: FF121211-04, FF010912-01, FF013112-01, FF020912-01, FF022212-03, FF022912-01, FF040412-01, FF051012-01
Quantity recalled41 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0288-2013
FDA event ID
63586
Recalling firm
Trivascular, Santa Rosa, CA
Classification
Class II
Status
Terminated
Recall initiated
19 Oct 2012
FDA classified
13 Nov 2012
Reported
21 Nov 2012
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Packaging Defects

Timeline

19 Oct 2012Recall initiated by company
13 Nov 2012Classified by FDA+25 days
21 Nov 2012Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Selected lots of the Fill Polymer Kit (European Model Number: TVFP14- B) in European distributor inventory only that potentially contain leaking stopcocks due to possible cracks in the stopcock component. TriVascular indicated that these Fill Polymer Kits are for use with the Ovation or Ovation Prime Abdominal Stent Graft Systems that may display leaks upon initiation of mix but prior to use in

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

TriVascular brand Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European Model Number: TVFP14-B, (in European distributor inventory only); PRODUCT Usage: The TriVascular Ovation Abdominal Stent Graft System is indicated in subjects diagnosed with an aneurysm in the abdominal aorta having vascular morphology suitable for endovascular repair, including: (1) Adequate iliac/femoral access compatible with vascular access techniques, devices, and/or accessories, (2) Non-aneurysmal proximal aortic neck: - with a length of at least 7 mm proximal to the aneurysm, - with an inner wall diameter of no less than 16 mm and no greater than 30 mm, and - with an aortic angle of d 60 degrees if proximal neck is e 10 mm and d 45 degrees if proximal neck is < 10 mm, (3) Non-aneurysmal distal iliac landing zone: - with a length of at least 10 mm, - with an inner wall diameter of no less than 8 mm and no greater than 20 mm

Where it was distributed

Worldwide Distribution - US Nationwide and the countries of Belgium, Cyprus, Italy, Spain, Switzerland and Turkey. TriVascular stated that no purchased devices are in hospital inventory in Europe or the U.S.

Nationwide

Stores named: Walmart

Questions about this recall

Is Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European recalled?

Yes. Trivascular recalled Fill Polymer Kit, Endovascular Graft, Aneurysm Treatment System, European on 19 Oct 2012. The recall is Class II and its status is Terminated. Recall number Z-0288-2013.

Why was it recalled?

Selected lots of the Fill Polymer Kit (European Model Number: TVFP14- B) in European distributor inventory only that potentially contain leaking stopcocks due to possible cracks in the stopcock component. TriVascular indicated that these Fill Polymer Kits are for use with the Ovation or Ovation Prime Abdominal Stent Graft Systems that may display leaks upon initiation of mix but prior to use in

Which lots are affected?

The following lot numbers are in distributor inventory in Europe: FF121211-04, FF010912-01, FF013112-01, FF020912-01, FF022212-03, FF022912-01, FF040412-01, FF051012-01

Where was it sold?

Worldwide Distribution - US Nationwide and the countries of Belgium, Cyprus, Italy, Spain, Switzerland and Turkey. TriVascular stated that no purchased devices are in hospital inventory in Europe or the U.S.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.