CLASS II Device recall · Reported 23 Sep 2015 · FDA Enforcement Report
Fluorescence Imaging Procedure Kits used recalled by Intuitive Surgical
Intuitive Surgical is recalling Fluorescence Imaging Procedure Kits used with the da Vinci Si and Xi Surgical Systems, distributed nationwide in the US. The recall was initiated on 9 Sep 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2782-2015
- FDA event ID
- 72158
- Recalling firm
- Intuitive Surgical, Sunnyvale, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 9 Sep 2015
- FDA classified
- 17 Sep 2015
- Reported
- 23 Sep 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Surgical Instruments
Timeline
| 9 Sep 2015 | Recall initiated by company | |
| 17 Sep 2015 | Classified by FDA | +8 days |
| 23 Sep 2015 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
B. Braun 2 gang 4-way stopcocks in the Fluorescence Imaging Procedure Kit may experience damage to the sterile packaging during shipping. The damage can result in tears to the packaging and has the potential to cause a breach in sterility of the supplied stopcocks.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Fluorescence Imaging Procedure Kits used with the da Vinci Si and Xi Surgical Systems; Model numbers: 950156-01, 950156-02, 950156-03. Product Usage: General and plastic surgery: The da Vinci Firefly Imaging system is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging.
Where it was distributed
US Nationwide Distribution
Questions about this recall
Is Fluorescence Imaging Procedure Kits used with the da Vinci Si and Xi Surgical Systems recalled?
Yes. Intuitive Surgical recalled Fluorescence Imaging Procedure Kits used with the da Vinci Si and Xi Surgical Systems on 9 Sep 2015. The recall is Class II and its status is Terminated. Recall number Z-2782-2015.
Why was it recalled?
B. Braun 2 gang 4-way stopcocks in the Fluorescence Imaging Procedure Kit may experience damage to the sterile packaging during shipping. The damage can result in tears to the packaging and has the potential to cause a breach in sterility of the supplied stopcocks.
Which lots are affected?
Product number: 950156-01: Lot numbers: NQ141201 NQ141502 NQ142201 NQ142901 NQ143201 NQ143301. Product number 950156-02: Lot numbers: NQ143501 NQ143502 NQ143601 NQ143602 NQ143803 NQ143901 NQ143902. Product number 950156-03: Lot numbers : NQ144001 NQ150501 NQ143902 NQ150601 NQ143901 NQ150602 NQ143803 NQ150703 NQ143601 NQ150701 NQ143602 NQ150702 NQ144502 NQ150801 NQ144801 NQ150901 NQ145202 NQ150802 NQ144901 NQ151001 NQ145301 NQ151002 NQ145201 NQ151101 NQ150301 NQ150501 NQ144001.
Where was it sold?
US Nationwide Distribution
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Intuitive Surgical
All 256| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | da Vinci SP Instrument Arm Drape, SP recalled by Intuitive Surgical | Intuitive Surgical | Other | Nationwide |
| Device | CLASS II | Da Vinci ASSY, PSS, SP1098 recalled by Intuitive Surgical | Intuitive Surgical | Device Malfunction | Nationwide |
| Device | CLASS II | Da Vinci 5 Assy,DV5 Tower,IS5000 recalled by Intuitive Surgical | Intuitive Surgical | Device Malfunction | 49 states |
| Device | CLASS I | Intuitive SureForm 30 Gray Reloads Reference Numbers recalled | Intuitive Surgical | Other | 49 states |
| Device | CLASS II | Da Vinci X and Da Vinci Xi da Vinci X, Xi Surgical System recalled | Intuitive Surgical | Device Malfunction | Nationwide |
Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Sterility | Nationwide |